Foot length, an accurate predictor of nasotracheal tube length in neonates

Abstract

BACKGROUND Existing guidelines for optimal positioning of endotracheal tubes in neonates are based on scanty data and relate to measurements that are either non-linear or poorly reproducible in sick infants. Foot length can be measured simply and rapidly and is related to a number of external body measurements.

OBJECTIVES To evaluate the relation of foot length to nasotracheal length in direct measurements at post mortem examinations, and then compare its clinical relevance with traditional weight based estimates in a randomised controlled trial.

METHODS The dimensions of the upper airway were measured at autopsy in 39 infants with median (range) postmenstrual age and birth weight of 32 (24–43) weeks and 1630 (640–3530) g. The regression equations with 95% prediction intervals were calculated to estimate the optimal nasotracheal length from foot length. In a randomised trial, 59 neonates were nasally intubated according to foot length and body weight based estimates to assess the achievement of “optimal” and “satisfactory” tube placements.

RESULTS In the direct measurements of the airway at autopsy, foot length was a better predictor of nasotracheal distances (r 2 = 0.79) than body weight, gestational age, and head circumference (r 2 = 0.67, 0.58, and 0.60 respectively). Measurement of foot length was easy and highly reproducible. In the randomised controlled trial, there were no significant differences between the foot length and body weight based estimates in the rates of optimal (44% v56%) and satisfactory (83% v 72%) endotracheal tube placements.

CONCLUSIONS Foot length is a reliable and reproducible predictor of nasotracheal tube length and is at least as accurate as the conventional weight based estimation. This method may be particularly valuable in sick unstable infants.

  • foot length
  • endotracheal tube

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Accurate positioning of the endotracheal tube (ETT) in mechanically ventilated neonates is essential for optimal ventilation, yet as many as half of intubated babies may have atelectasis secondary to endobronchial placement of tubes.1 Indeed, malposition of the ETT is the most common unpredicted radiographic finding necessitating intervention.2 The accepted optimum position for the tip of the ETT is in the midtrachea, and a number of nomograms are available to predict this depth from external body measurements such as weight, head circumference, crown-heel or crown-rump length.1 ,3-5 However, many of these studies have their limitations. The original cadaveric work3 included only 15 infants, only one of these being less than 1000 g in weight. Today, a large number of infants admitted to the neonatal intensive care unit weigh less than 1000 g. Although studies show crown-rump5or crown-heel lengths6 to be the best predictors of midtracheal distance, they are difficult to measure accurately in sick babies and lack reproducibility. Weight is easy to obtain accurately but it is a non-linear measurement, and predictions based on it may be misleading in oedematous or growth retarded infants. Head circumference is related more to brain mass than body length and may be misleading in infants with cephalohaematoma, caput succedaneum, and microcephaly or macrocephaly.

It has been shown that the birth weights and crown-rump and crown-heel lengths of neonates, particularly premature babies, can be accurately estimated from measurement of their foot lengths, and that the latter measurement can be made simply, rapidly, and safely even in critically ill neonates.7 We hypothesised that foot length may provide a simple and accurate prediction of optimal ETT length. We evaluated the usefulness of this measurement in estimating the naris to midtracheal length in neonates during direct measurements of the upper airway at autopsy, and have tested its clinical relevance against the traditional weight derived estimates in a randomised clinical trial.

Patients and methods

OVERVIEW

We measured the naris to midtracheal length at autopsy in 39 neonatal deaths and non-macerated still births, and investigated its relation to external body measurements (post mortem study). Foot length was the best variable for prediction of naris to midtracheal length. We then compared the accuracy and reliability of the estimation of ETT length from foot length against the body weight in a prospective randomised controlled trial.

POST MORTEM STUDY

Data were collected on the dimensions of the upper airway and external body measurements at autopsy in 39 neonatal deaths and non-macerated fresh stillbirths; the parents had given consent for post mortem examination. We excluded babies with fetal hydrops, congenital malformations, or chromosomal anomalies. Post mortem body weight, occipitofrontal head circumference (OFC), crown-rump length (CRL), shoulder-umbilicus length (SUL), sternal notch-umbilicus length (STUL), and foot length were measured before autopsy dissection.

The upper airway dimensions were measured with the trachea in situ. After removal of the sternum, the left lung and heart were displaced to the right. The carina was identified and incised. A firm graduated 3 mm catheter was then inserted and passed upwards through the trachea, larynx, pharynx, and nasal cavity until the tip of the tube appeared at the nostril. The nose-carina (NC) length was then measured using the gradations on the tube. The vocal cord-carina (VC) distance was measured directly after removal of the thoracic organs and opening of the posterior wall of the larynx and trachea. The naris-midtracheal length (NMTL) was calculated as NC−(0.5 × VC). All these lengths were measured in mm.

Statistical analysis

Linear regression analysis was used to investigate the relation of NMTL to potential predictors, including corrected gestational age, weight at autopsy, OFC, CRL, SUL, STUL, and foot length.

CLINICAL STUDY

Variability in foot length measurement

Interobserver variability in foot length measurement was assessed by having the same foot measured by 10 different nurses within six hours in seven infants. Intraobserver variability was evaluated by measurement of the same foot by the same observer on 10 occasions in four babies. The within subject standard deviations (ζω) were estimated by one way analysis of variance, and the repeatability (the difference between two measurements for the same subject for 95% pairs of observations) was calculated as 2.77ζω.

We designed a prospective randomised clinical study to assess the accuracy of the new foot length based prediction against the traditional weight based estimates. In view of the nature of the study, local research ethics committees did not stipulate informed parental consent while granting approval for the study.

Study subjects

The study subjects were all neonates needing nasotracheal intubation by the neonatal unit staff, including those intubated by the neonatal interhospital transport team. We excluded babies who were hydropic, dysmorphic, or intubated solely by the referring paediatricians.

Study design

Randomisation was performed in advance with computer generated codes, and assignments were inserted in sealed consecutively numbered opaque envelopes. The randomisation code was revealed when an eligible infant was considered for nasotracheal intubation. If nasotracheal intubation was unsuccessful, then the envelope was discarded.

In infants assigned to “weight”, the birth weight or the most recent weight (if greater than the birth weight) were used to determine the predicted nasotracheal tube length. For infants drawing “foot length”, this was defined as the distance from the tip of the heel to the tip of the great toe, and this length was measured using a plastic ruler. The foot length was rounded to the nearest 0.5 cm mark. The estimated ETT length was then read off the appropriate charts. For both groups, up to 0.5 cm was added to the predicted ETT length to allow for the fixation of the tube. The infant was then nasally intubated using the ETT appropriate to the infant's weight and naris size.

The nasotracheal tubes were fixed using the method described by Gregory.8 The equality of air entry was checked by auscultation, and radiological confirmation of the tip of the ETT was obtained as soon as possible. The chest radiograph after intubation was obtained with the infant lying directly on the roentgenogram plate with the head in the midline and the neck in the neutral position. The beam was centred on the sternal angle.

Outcome measures

The primary outcome measure was the achievement of “optimal midtracheal” ETT position defined as the tip of the tube at the body of the first thoracic vertebra.9 Secondary outcome measures were (a) achievement of “satisfactory” ETT position defined as the tip of the tube located within the trachea from between the medial ends of the clavicle to at least 1.5 cm above the carina, and (b) the need for reintubation. If the ETT position was considered unsatisfactory, needing adjustment or reintubation, this was considered a failure.

Statistical analysis

Based on previous data, we assumed that body weight based estimates will achieve optimal ETT placement in 60%. To enable a 30% difference in the achievement of optimal ETT placement with 80% power at the 5% significance level, we needed 32 infants in each group. Statistical analysis was performed with STATA 5.0 for Macintosh (Stata Corporation, College Station, Texas, USA). The basic characteristics of the two groups were compared by Mann-Whitney U test for continuous and χ2 test for the categorical variables. The outcome data were compared by the two tailed χ2 test with Fisher's exact modification where necessary.

Results

POST MORTEM STUDY

The median (range) postmenstrual age and birth weight of the infants were 32 (24–43) weeks and 1630 (640–3530) g respectively. Ten infants were less than 30 weeks gestation, and seven weighed less than 1000 g. Table 1 shows the regression equations and residual SDs for each of the anthropometric measurements (weight, OFC, CRL, SUL, and foot length) and gestational age with NMTL. Foot length was the best single anthropometric measure for predicting the NMTL as well as other upper airway dimensions such as NC and VC (the regression data for the latter not shown) followed closely by the CRL. Figure 1 shows the graph for estimating the optimal ETT length from foot length with 95% prediction intervals. We then used this graph to estimate nasotracheal tube length for intubation in the randomised trial.

Table 1

Regression equations for naris to midtracheal distance in mm derived from gestational age (Gest), body weight (Wt), Occipitofrontal head circumference (OFC), crown-rump length (CRL), shoulder to umbilicus length (SUL), and foot length (FL)

Figure 1

Graph for prediction of optimal endotracheal tube (ETT) length (with 95% prediction intervals) for nasotracheal intubation based on measurement of foot length. (Add 0.5 cm to the measured ETT length if the tube is sutured to a tape.)

CLINICAL STUDY

Measurement error

The within subject standard deviation in the interobserver study was 1.47 mm, giving a repeatability of 4 mm. For the intraobserver variation, the within subject standard deviation was 0.9 mm with a repeatability of 2.5 mm.

Randomised controlled trial of nasotracheal intubation in neonates

Of the 72 infants randomised in the trial, 13 could not be successfully nasally intubated. Confirmatory chest radiographs after intubation were unavailable in a further four (two in each group) infants, although all these infants were successfully ventilated and none required reintubation. The baseline characteristics were similar for the two groups of randomised infants (table 2). Seventeen infants weighed less than 1000 g at intubation.

Table 2

Baseline characteristics of study infants

There were no statistically significant differences in the rates of achievement of optimal and satisfactory placements of the ETTs between the two groups (table 3). In three infants from the body weight group and two from the foot length group, the ETTs needed repositioning. Of 10 infants in whom the ETT tip was located at vertebral bodies C6–8, one was between, and nine were above, the medial ends of the clavicles. Of 32 infants in whom the ETT tip was between the first and second thoracic vertebrae, 19 were between the medial ends of the clavicles, with the remainder just below it. Of the five ETT tips at T4–5, two were in the low trachea (but safe) and three were at or below the carina.

Table 3

Endotracheal tube (ETT) placements

Discussion

Our direct measurements of the airway dimensions in post mortem examinations showed that foot length was the best predictor of naris-midtracheal length. In a subsequent randomised clinical trial against traditional body weight based estimates, foot length proved an accurate, reproducible, and easy to measure alternative predictor for estimating optimal nasotracheal tube length for neonatal intubation.

STRENGTHS

Our hypothesis that foot length provides an accurate prediction of naris-midtracheal length in babies was confirmed by direct measurements of the airway at autopsy. We excluded dysmorphic infants from both studies as the tracheal length in such infants may be shorter than normal.10 We also precisely defined the optimum position for the tip of the ETT in the midtrachea as corresponding radiologically to the first thoracic vertebra.9 We included a pragmatic measure of satisfactory ETT length because babies who may not have an ideally placed ETT can still be successfully ventilated. Flexion of the head decreases the naris to carina distance (a risk for bronchial intubation) by up to 1 cm in infants under 1000 g and up to 1.6 cm in bigger infants.5 We therefore chose the distance of 1.5 cm above the carina as a safe position. Although a number of measurements other than body weight have been suggested for estimation of optimal depth of ETT placement in babies,6 ,11-15 they have not been evaluated in randomised controlled trials.

LIMITATIONS

We based the predictive equations on the data collected at autopsy. In vivo traction from the diaphragm increases the length of the trachea during inspiration. Loss of this traction effect after death results in a shorter trachea. We attempted to limit this effect by measuring tracheal lengths in situ before its attachments were severed. Observations at bronchoscopy in live infants,16and radiological evaluations6 ,12 have been used to provide in vivo airway dimensions, but these studies have not included small premature infants limiting their application to current neonatal practice.

We did not use a specially designed foot gauge as used by James and others,7 but found very little intraobserver and interobserver variability in the measurement of foot length using simple plastic rulers. We rounded off the foot length measurement to the nearest 0.5 cm mark, as the repeatability of this measurement in clinical practice was 4 mm. As our units use the nasotracheal route for babies requiring medium to long term ventilation, we calculated the nasotracheal tube lengths. Our results are not applicable to the oral route.

STRENGTHS AND WEAKNESSES IN RELATION TO OTHER STUDIES

Although attempts have been made in recent years to locate the ETT tip sonographically at the time of intubation17 or by using a fibre optic light18 or magnetic ring embedded in the ETT tube (TRACH MATE),19 such equipment is either not widely available or adds complexity. Optimal ETT length is therefore commonly estimated from nomograms based on some external body measurement. The traditionally used weight or head circumference based graphs are derived from anatomic measurements of excised tissues at autopsies of mainly term infants,3 ,4 limiting their relevance to small preterm babies. Although Rotschild and colleagues5 provided regression equations for optimal ETT lengths, these require measurements of crown-heel and crown-rump lengths, which are difficult to obtain accurately and without disturbing a critically ill ventilated baby. Shukla and colleagues13 suggested a rapid method for estimating orotracheal and nasotracheal tube lengths from nasal-tragus and sternal lengths. In a prospective study, they reported achievement of a perfect ETT tube position in 48 of 50 infants using their predictors. However, this was a non-randomised study and no definition of perfect ETT position was provided. Our study shows that, despite their excellent performance in post mortem studies, predictors of optimal ETT length achieve such positions in only half of the clinical intubations, with a fifth having unsatisfactory tube placements. Hence, a rigorous controlled clinical evaluation of newer methods of estimating optimal ETT placement is necessary.

EVALUATION OF THE STUDY: POSSIBLE MECHANISMS AND IMPLICATIONS FOR CLINICIANS

Foot length has been shown to be related positively to gestational age20 ,21 and length in fetuses,22 and various indices of body size in both large for dates and appropriately grown babies of all gestational ages between 26 and 42 weeks.7 In the latter study,7 there were particularly pronounced correlations of foot length with body weight (r = 0.95) and crown-heel length (r = 0.96) in premature babies. As tracheal length is a linear body measurement, it is not surprising that it correlates more closely with linear measures such as foot length or crown-rump length than body weight or head circumference.

Unlike the illuminated or magnetic ring embedded ETTs, use of foot length for prediction of ETT length does require radiographic confirmation of the tube position, necessitating exposure to ionising radiation every time the location of the tube is called into question. Measurement of foot length, however, does not require any additional equipment apart from a widely available transparent plastic ruler, and can be performed easily and reproducibly even in very unstable babies. Our clinical trial shows that foot length provides an alternative, reliable, and accurate method for determining the optimal ETT length, particularly in sick unstable infants.