Reproducibility of the cervical range of motion (CROM) device for individuals with sub-acute whiplash associated disorders

Eur Spine J. 2012 May;21(5):872-8. doi: 10.1007/s00586-011-2096-8. Epub 2011 Dec 3.

Abstract

Purpose: The objective of this study was to evaluate the reproducibility of the cervical range of motion device when measuring both active and passive range of motion in a group of individuals with sub-acute Whiplash Associated Disorders (WAD).

Methods: Participants were recruited as part of a large multi-centre Randomised Controlled Trial from UK emergency departments. Experienced research physiotherapists measured active and passive cervical spine movements in all directions. Both intra- and inter-observer reliability and agreement were assessed using the intra-class correlation coefficient, standard error of measurement and limits of agreement methods.

Results: Different groups of 39 and 19 subjects were included in the intra and inter-observer studies, respectively. The CROM device demonstrated substantial intra- and inter-observer reliability and agreement for all the active and passive half-cycle movements (ICC range 0.82-0.99) with the exception of one (passive right lateral flexion for inter-observer; ICC 0.77).

Conclusions: The CROM device has proven to be a reproducible measurement method for a symptomatic WAD population using the measurement protocol described and can be used with confidence to differentiate individuals according to a single measurement.

Publication types

  • Validation Study

MeSH terms

  • Adult
  • Arthrometry, Articular / instrumentation*
  • Arthrometry, Articular / methods
  • Cervical Vertebrae / physiopathology*
  • Humans
  • Male
  • Middle Aged
  • Observer Variation
  • Range of Motion, Articular / physiology*
  • Reproducibility of Results
  • Retrospective Studies
  • Whiplash Injuries / diagnosis
  • Whiplash Injuries / physiopathology*