Is visual assessment of jaundice reliable as a screening tool to detect significant neonatal hyperbilirubinemia?

J Pediatr. 2008 Jun;152(6):782-7, 787.e1-2. doi: 10.1016/j.jpeds.2007.11.003. Epub 2008 Jan 22.

Abstract

Objective: To assess the reliability of visual assessment of bilirubin levels (BiliEye) in newborns as a screening tool to detect significant neonatal hyperbilirubinemia.

Study design: 5 neonatologists and 17 nurses estimated 3,532 BiliEye in 1,129 term and late preterm (> or = 35 weeks) infants before discharge from the nursery, at 62 +/- 24 hours. Total serum bilirubin (TSB) levels were measured concomitantly.

Results: Mean TSB and BiliEye were 6.7 +/- 2.9 mg/dL (range, 0.4-18.2 mg/dL) and 6.6 +/- 3.2 mg/dL (range, 0.0-17.2 mg/dL), respectively, with good correlation (Pearson's r = 0.752, P < .0001), but other measures of agreement were poor. 61.5% of the 109 babies with TSB levels in high-risk zones were clinically misclassified. The area under curve (AUC) of the receiver-operating characteristics plotted for these high-risk zones was 0.825, but became low for early discharge (< or = 36 hours; AUC = 0.638) and late preterm (35-37 weeks; AUC = 0.613). There was significant interobserver variation (low weighted kappa, 0.363).

Conclusions: Although there was good correlation between BiliEye and actual TSB level, visual assessment was unreliable as a screening tool to detect significant neonatal hyperbilirubinemia before discharge. Babies with TSB levels within high-risk zones may be clinically misdiagnosed as low-risk, resulting in inadequate follow-up.

MeSH terms

  • Adult
  • Bilirubin / blood
  • Female
  • Humans
  • Infant, Newborn
  • Jaundice, Neonatal / blood
  • Jaundice, Neonatal / diagnosis*
  • Male
  • Neonatal Screening / methods
  • Physical Examination / methods
  • Reproducibility of Results