Original Contribution
Quantification of Muscle Volume by Echography: Comparison with MRI Data on Subjects in Long-Term Bed Rest

https://doi.org/10.1016/j.ultrasmedbio.2009.01.004Get rights and content

Abstract

The objective of this study was to determine the accuracy of an echographic method for measuring the change in leg muscle volume against the gold standard, magnetic resonance imaging (MRI). Leg muscle volume was measured using an echographic scanner, which consisted of two metallic rails on which a probe holder moved via an electric engine. Ten to 20 transverse muscle views were collected along the area scanned, and the muscle cross-sectional area (CSA, cm2) was measured on each of them. The integration of all the CSAs along the scanned area provided the muscle volume (cm3). Echographic results were compared with MRI data on 24 subjects undergoing 60 d of bed rest (8 control “Con,” 8 with exercise countermeasures “Ex” and 8 with nutrition countermeasures “Nut”). The vastus intermedius (VI) and the vastus medialis (VM) volumes decreased significantly and similarly in both Con and Nut (VI, –17%; VM, –21%; p < 0.02). In the Ex group, the VI and VM did not change significantly. The correlation coefficient between the muscle volume change measured with the echographic and MRI methods was 0.78. The present study confirms that the echographic scanner is sufficiently accurate for assessing muscle volume changes and detects the effect of exercise countermeasures on muscle volume during long-term bed rest. (E-mail: [email protected])

Introduction

Presently, muscle volume in humans is estimated by relatively simple and inaccurate anthropometric measurements (e.g., measuring the thigh perimeter at different levels) or with sophisticated, cumbersome and expensive magnetic resonance imaging (MRI) (Trappe et al. 2001b) or computed tomography (Radegran et al. 1999) methodologies. Quantifying muscle volume loss is critical for the treatment of sarcopenia (the age-related loss of muscle mass) and for assessing the recovery of muscle atrophy induced by immobilization after trauma or surgery. Accurate measurements of muscle volume are also important for humans travelling on long-duration space missions, such as on the International Space Station or future trips to the moon or other planets with reduced gravity. Our objective was to develop a new tool for assessing physiological and pathological changes in muscle volume that was accurate and could be used in the aforementioned populations more feasibly than the current approaches.

Because of its size, low cost and relative ease of use, echography could serve as such a tool to accurately measure human skeletal muscle volume. However, 3-D echographs do not have a sufficient scanning angle for insonating and measuring large parts of human muscles (i.e., muscle volume). Thus, we developed a mechanical probe holder that could move a 2-D probe along the muscles and measure muscle volume by scanning a dedicated part of the muscle. The system was validated on subjects in which muscle atrophy was induced by putting them in head-down bed rest (HDBR) for two months. The study was designed to mimic the undesirable effects of microgravity on the main physiological systems, including the cardiovascular and musculoskeletal systems, and to test exercise and nutritional countermeasures designed to prevent these effects.

Loss of muscle mass and function has been one of the hallmarks of exposure to real and simulated microgravity (Adams et al., 2003, Alkner and Tesch, 2004a, Alkner and Tesch, 2004b, Fitts et al., 2000, Fitts et al., 2001, Tesch et al., 2005, Trappe et al., 2001a). Muscle volume of the knee extensors decreases significantly after short and long-duration space flights (8 to 17 d: 5 to 15%; 112 to 196 d: 10%), despite exercise countermeasures and dietary control (caloric intake requirements). However, because of technical constraints, measurements (usually MRI) are restricted to before and after flight, with no “real time” indication of the efficacy of the prescribed countermeasures (LeBlanc et al. 2000). Ground-based studies show the knee extensor muscle volume loss approaches 20% over several months and can be partially or completely prevented with exercise countermeasures (Akima et al., 2001, Alkner and Tesch, 2004a, Alkner and Tesch, 2004b, Blottner et al., 2006, Gallagher et al., 2005, Tesch and Trieschmann, 2004, Trappe et al., 2004, Trappe et al., 2007).

The objective of this study was to determine the accuracy of an echographic method for measuring the change in leg muscle volume against the gold standard, MRI. Echographic and MRI measurements were performed on the same subjects before and at the end of two months of HDBR. Some of the subjects used exercise or nutritional countermeasures designed to reduce or prevent muscle atrophy.

Section snippets

Principle of the ultrasound muscle scanner

A conventional echographic probe (5 to 10 MHz sector array, Challenge 2000, Esaote, Firenze, Italy) was fixed on a probe holder fabricated by the investigators that moved continuously along two parallel metallic rails (25 cm long and 6 cm apart) via a small electric engine (Model: ESCAP, reference:26N58-216E1, brushed DC, 12v, 5.7w; by PORTESCAP Cie, La Chaux-de-Fonds Canton de Fribourg, Switzerland) (Fig. 1). The DC motor uses a rotor, of which the active part simply consists of a

Results

The system was tested in vitro on different soft materials (silicone cylinders and cubes, and pieces of beefsteak), and the error on the volume (100 to 500 cm3) was less than 3%. The volume of each piece was determined by measuring the water displacement caused by immersion and compared with the volume measured on the echographic views collected with the ultrasound scanner.

The VI and VM were investigated with the lowest part of the probe support in contact with the upper pole of the patella,

Discussion

The light probe holder (300 g) was easy to place and maintain on the thigh of the subjects during the scanning operation. One or two runs of the motorized probe holder were performed at high speed (5 cm/s) to confirm the muscle of interest was scanned appropriately, and one acquisition run (1.5 cm/s) was sufficient to collect the images for the muscle volume calculation. Automatic muscle edge detection significantly shortened the time required for measuring all of the muscle CSAs and ultimately

Conclusion

The muscle volume data measured by the echographic scanner confirmed the MRI data. That is, the echographic method tracked the muscle volume loss because of the bed rest, the prevention of muscle loss with the exercise countermeasure or the lack of influence of the nutrition countermeasure. The ultrasound scanner allowed the accurate evaluation of the volume change of various muscles by an easy and semi-automated method.

Acknowledgments

The study WISE-2005 was sponsored by the European Space Agency (ESA), the National Aeronautics and Space Administration of the USA (NASA), the Canadian Space Agency (CSA) and the French space agency “Centre National d'Études Spatiales” (CNES). The authors would like to thank the radiology staff at Rangueil Hospital (Virginie Lorite, Ghylene Enrique, Christine Mazars, Muriel Rogawski) for their efforts with this investigation and the sonographers of the Dept Med Nucl et Ultrasons (M. Porcher, V.

References (17)

  • G.R. Adams et al.

    Skeletal muscle unweighting: Spaceflight and ground-based models

    J Appl Physiol

    (2003)
  • H. Akima et al.

    Inactivity and muscle: Effect of resistance training during bed rest on muscle size in the lower limb

    Acta Physiol Scand

    (2001)
  • B.A. Alkner et al.

    Efficacy of a gravity-independent resistance exercise device as a countermeasure to muscle atrophy during 29-day bed rest

    Acta Physiol Scand

    (2004)
  • B.A. Alkner et al.

    Knee extensor and plantar flexor muscle size and function following 90 days of bed rest with or without resistance exercise

    Eur J Appl Physiol

    (2004)
  • D. Blottner et al.

    Human skeletal muscle structure and function preserved by vibration muscle exercise following 55 days of bed rest

    Eur J Appl Physiol

    (2006)
  • R.H. Fitts et al.

    Physiology of a microgravity environment. Invited review: Microgravity and skeletal muscle

    J Appl Physiol

    (2000)
  • R.H. Fitts et al.

    Functional and structural adaptations of skeletal muscle to microgravity

    J Exper Biol

    (2001)
  • P. Gallagher et al.

    Effects of 84-days of bedrest and resistance training on single muscle fibre myosin heavy chain distribution in human vastus lateralis and soleus muscles

    Acta Physiol Scand

    (2005)
There are more references available in the full text version of this article.

Cited by (35)

  • Supplemental parenteral nutrition improves immunity with unchanged carbohydrate and protein metabolism in critically ill patients: The SPN2 randomized tracer study

    2019, Clinical Nutrition
    Citation Excerpt :

    Infectious complications were collected prospectively respecting the definitions of the 2005 consensus conference [9]. Muscle mass was investigated using the panoramic brightness-mode ultrasound determination [10] of the cross-sectional area (CSA) of the quadriceps femoris [11] on days 4, 9–10, and 15. A custom harness was fixed on the patient's thigh to ensure fixed distances between the base of the patella, and the level of muscle measurement.

  • Reliability of ultrasonographic arm muscle thickness measurement by various levels of health care providers in ICU

    2018, Clinical Nutrition ESPEN
    Citation Excerpt :

    It's non-invasive nature, bedside availability, repeatability over multiple time with zero adverse effect makes it a tool of choice among critical care physicians. Also, its accuracy is comparable with computerized tomographic (CT) scan or magnetic resonance imaging (MRI) [8,10]. USG has been shown to be very useful tool in the hands of critical care physicians as shown by excellent intra- and inter-observer reliability with minimum variability for measurement of muscle thickness [11–14].

  • Reliability of bedside ultrasound for measurement of quadriceps muscle thickness in critically ill patients with acute kidney injury

    2017, Clinical Nutrition
    Citation Excerpt :

    Ultrasonography (US) is widely available and easily applicable at the bedside for different skeletal muscle groups, even in critically ill patients [3]. It allows the detection of muscle wasting during ICU stay [4], and seems as accurate as CT or MRI in measuring muscle mass [5]. No data are currently available on the reliability of US measurements in patients with AKI.

View all citing articles on Scopus
View full text