Introduction

Since its introduction, CT has become the major imaging modality for investigating chest disorders, a position reinforced over years by the various technological advances successively introduced in clinical practice. Among them, multidetector CT (MDCT) has definitely influenced the evaluation of airways and lung diseases, allowing examination of the entire thorax without gaps, providing simultaneous availability of several series of images from a single data set with a dramatic reduction in breathing artefacts inherent to the high speed of data acquisitions [13]. This technology has also become the gold standard for diagnosing pulmonary vascular diseases, combining high spatial resolution and an optimal enhancement of thoracic circulations, exemplified in the diagnosis of acute pulmonary embolism [4]. More recently, it has become possible to integrate cardiac and coronary artery imaging into chest CT examinations, a concept applicable to numerous thoracic diseases [5]. The consequence of such developments is the substantial increase in the radiation dose delivered to patients, individually but also collectively [6], and the subsequent risks, even small, of cancer induction by low-level radiation [7, 8]. Facing the challenge of providing diagnostic image quality at the lowest radiation dose, the radiological community has modified chest CT protocols at the pace of new technology implementations aiming to reduce radiation exposure, such as anatomical tube current modulation, ECG-controlled tube current modulation or dynamically adjustable pre-patient collimation of the X-ray beam in the z-axis direction [9, 10]. Simple dose saving behaviours have also been introduced in the radiologists’ daily practice, such as the adjustment of the CT parameters selected at the console, further optimized when individually-adapted, which can be associated with automatic tube current modulation systems [11, 12]. More recently, another option to save dose is the use of iterative reconstruction techniques [13].This two-part study was initiated to evaluate the performance of a recently introduced iterative reconstruction technique (Iterative Reconstruction in Image Space IRIS) [14]. In part 1 of the study, we evaluated image noise reduction achievable on routine weight-based chest CT examinations; in part 2, we implemented this technique for saving dose in a nonselected adult population.

Materials and methods

Study population

In order to evaluate the level of noise reduction and overall image quality with an iterative reconstruction algorithm, chest CT examinations of 32 consecutive patients referred for a standard CT angiographic examination were included, each data set being systematically reconstructed with regular filtered back-projection (FBP), which is the standard approach for medical CT today, and using an iterative reconstruction technique (IRIS; Siemens Healthcare, Forchheim, Germany). The sample size necessary for this evaluation was estimated on the basis of the following parameters: (a) the main purpose of this study was to compare the image noise level between two techniques, namely the standard FBP reconstructions and the iterative reconstructions obtained after 3 iterations; (b) on the standard FBP reconstructions, the mean image noise measured at the level of the trachea on mediastinal images had been previously estimated at 22.6 ± 6 Hounsfield units (HU) [15]; (c) our hypothesis being a noise reduction of 20% with the iterative reconstruction technique using 3 iterations, and assuming a correlation value of 0.2 between two measurements of noise with the two reconstruction techniques, an estimated number of 32 patients was needed for 90% power with significance at 5%.

The study protocol was approved by our institutional Ethics Committee; no patient’s informed consent was required to generate the two series of reconstructions; the retrospective analysis of data was possible with waiver of patients’ informed consent. The study group was composed of 19 males and 13 females with a mean (±SD) age of 58 (±18) years (range: 23-91), referred for various respiratory disorders, including bronchial and/or lung parenchymal (n = 19), pulmonary vascular (n = 8), pleural (n = 3) and mediastinal (n = 2) diseases. The mean (± SD) weight and the mean (±SD) body mass index of our study group were 66.1 (±12.5) kg (range: 50-100) and 23.5 (±5.5) kg/m2 (range: 17–42).

CT protocol

CT angiograms were obtained on a dual-source 128-slice MDCT system (Somatom Definition Flash, Siemens, Germany), using two tubes with single energy in a non electrocardiogram (ECG)-gated mode, with the following parameters: collimation: 64 × 2 × 0.6 mm with z-flying focal spot; weight-adapted selection of the kilovoltage for both tubes (ranging between 100 and 120 kV) with adapted milliamperage setting (ranging between 90 and 100 effective mAs); 4D dose modulation (Care Dose 4D); pitch: 3; rotation time: 0.28 s. The injection protocol consisted of the administration of 120 ml of a contrast agent with 350 mg of iodine per milliliter (Xenetix 350, Guerbet) at a flow rate of 4 ml/s. The examination was initiated by bolus tracking within the ascending aorta with a threshold of 150 HU to trigger data acquisition. The mean (±SD) dose-length-product of the CT angiograms was 125.5 (±53.5) mGy.cm, ranging from 48 to 263 mGy.cm. The average effective dose was retrospectively calculated by multiplying the DLP value by a conversion factor of 0.017 [16], leading to a mean value of 2.13 (±0.9) mSv (range: 0.81–4.47)

Image reconstruction

While increased spatial resolution is directly correlated with increased image noise in standard filtered back-projection reconstructions as they are used in CT practice today, iterative reconstruction approaches to a certain extent allow decoupling of spatial resolution and image noise [13]. In an iterative reconstruction, a correction loop is introduced. Once an image has been reconstructed from the measured projections, a ray-tracing in the image is performed to calculate new projections that exactly represent the reconstructed image. This step, called reprojection, simulates the CT measurement process with the image as measurement object. The deviation between measured and calculated projections is used to derive correction projections, reconstruct a correction image and update the original image. This loop is continued until the deviation between measured and calculated projections is smaller than a predefined limit. Each time the original image is updated, non-linear image processing algorithms are used to enhance spatial resolution at higher object contrasts and to reduce image noise in low contrast areas. This “regularization” step is essential for the noise reduction properties of an iterative reconstruction, whereas the repeated calculation of correction projections removes image artifacts introduced by the approximative nature of the filtered backprojection reconstruction. Regularization can as well be applied to the image data in an iterative loop without reprojection and calculation of correction projections [14]. In the “IRIS” approach, an iterative series of 3-dimensional non-linear image processing steps, corresponding to the regularization in a standard iterative reconstruction, is performed after reconstruction of an initial high resolution image. This high resolution image contains all measured information that is otherwise partially suppressed in a standard CT-reconstruction to obtain acceptable image noise levels. The non-linear image processing steps are locally adapted, according to the local image noise and image structure. They aim at maintaining or even enhancing spatial resolution at higher object contrasts, while reducing image noise in low contrast areas without degrading the image texture.

From each data set, 2 series of images (i.e., 1-mm thick images) were systematically generated on a prototype workstation: (a) lung and mediastinal images reconstructed with a standard FBP algorithm (Group 1) using a high spatial resolution kernel (B50) and a soft tissue kernel (B20), respectively; (b) lung and mediastinal images reconstructed with an iterative reconstruction technique (IRIS algorithm; Siemens) (Group 2) using a high spatial resolution iterative kernel (I50) and a soft tissue iterative kernel (I20), respectively. The iterative reconstruction kernels I50f and I20f are designed to closely match the spatial resolution (in terms of the 50% and 10% value of the modulation transfer function for high contrast objects) of the corresponding filtered back-projection kernels B50f and B20f, respectively. The assessment of the IRIS technique was made on 2 levels of iterations, including reconstructions using 3 iterations (Group 2a) and reconstructions using 5 iterations (Group 2b). All images were viewed at standard mediastinal (window width, 400 HU, window center, 40 HU) and lung parenchymal (window width, 1,600 HU; window center, -600 HU) window settings.

CT parameters evaluated

Image noise

The objective assessment of noise on Group 1, Group 2a and Group 2b images was obtained by measuring the standard deviation of pixel values in homogeneous regions-of-interest on mediastinal (B20 and I20) images (at two anatomical levels: the tracheal lumen above the aortic arch; the descending aorta at the level of the ventricular cavities) and on lung (B50 and I50) images (at one anatomical level: the tracheal lumen above the aortic arch) using the following methodology. On standard window settings, a circular region of interest (ROI) was positioned within the tracheal lumen or the aorta. The mean density value inside the ROI was chosen as the new window centre whereas the window width was arbitrarily defined between 100 HU and 200 HU, thus enabling the readers to confirm the homogeneity of noise within the tracheal or aortic lumina. On images viewed with the newly defined window settings, a circular ROI was again drawn within the trachea or the aorta to assess noise objectively. Care was taken to avoid superimposition of the ROI on the inner portion of the tracheal or aortic wall. The standard deviation of the measured density in HU defined the objective noise.

The visual perception of noise, defined by the grainy appearance of the CT images, was evaluated on mediastinal and lung images of Group 1, Group 2a and Group 2b images viewed with standard window settings. On each series of mediastinal and lung images, the image noise was rated as minimal (score 1) or moderate (score 2), by comparison with reference images (score 1 and 2 noise levels did not alter the identification of normal and/or abnormal structures), or important (score 3) when the image noise altered the identification of normal and/or abnormal structures.

Overall image quality

The overall image quality of lung and mediastinal images of each group of reconstructions was rated using a three-point scale. Reconstructions with distinct anatomic detail, no noise or minimal image noise were individually rated with a score 1 (excellent image quality). Reconstructions with clear anatomic detail and moderate increase in noise that did not affect their diagnostic value were individually rated with a score of 2 (good image quality without impairment of diagnostic accuracy), whereas examinations with a marked increase in noise altering their diagnostic value were individually rated with a score of 3 (non diagnostic examinations). When lung and mediastinal images of a given examination had different scores, the worst image quality was systematically considered to rate the overall image quality of the examination.

Lesion conspicuity in group 2

In order to evaluate the potential impact of iterative reconstructions on lesion conspicuity, we selected 3 lung images per patient on Group 1 reconstructions, each of them showing at least one of the following elementary lesions: ground-glass opacities, ill-defined micronodules, well-defined micronodules, septal lines, nonseptal lines, focal areas of hypoattenuation with (i.e., lung cysts) or without (i.e., centrilobular emphysematous lesions) definable walls. Lung images at the corresponding anatomical levels on Groups 2a and 2b reconstructions were selected in order to visually assess the presence of the elementary lesion(s) and to grade their conspicuity as equivalent, superior or inferior to that on Group 1 images. An additional evaluation of lesion conspicuity was made between Group 2a and Group 2b to determine a potential difference between the two series of images.

Conditions of image analysis

The image quality assessment was performed by 2 subspecialty thoracic radiologists (**BLINDED**) with 2 and 6 years experience in CT, respectively. They evaluated the CT parameters by consensus on a workstation. The FBP reconstructions were performed using the CT system’s built in reconstruction computer. For the iterative reconstructions, raw data were transferred to an offline PC provided by the vendor. The reconstructed images were dicomized and sent back to a clinical workstation for viewing and measurement of image noise.

Statistical analysis

Results were expressed as means and standard deviations for continuous variables, frequencies and percentages for categorical variables. For continuous variables, comparisons between reconstruction techniques were performed using a two-way analysis of variance (ANOVA). We used a linear mixed model with the patients considered as a random effect and the reconstruction technique as a fixed effect. For categorical parameters, comparisons were performed using the McNemar chi-square test. All pairwise comparisons were performed with a Bonferroni correction. The statistical significance was defined as p < 0.05. The statistical analyses were performed using the SAS software (SAS Institute Inc Cary, NC 25513,USA).

Results

Evaluation of image noise

Objective image noise

There was no statistically significant difference between the mean surfaces of the circular ROIs used for objective noise assessment (p > 0.05).

In Table 1,s the mean values of objective noise measurements on mediastinal and lung images reconstructed with FBP (Group 1), 3 iterations (Group 2a) and 5 iterations (Group 2b) are compared. There was a significant reduction in the objective noise in Group 2a compared with Group 1, with a mean noise reduction of 34.3% on lung images and a mean reduction of 31.5% at the level of the aorta and 37.2% at the level of the trachea on mediastinal images (ANOVA with Bonferroni correction; p < 0.0001).

Table 1 Comparison of image quality between iterative reconstructions and filtered back projections

The objective noise was further reduced on Group 2b images: (a) compared with Group 2a, the mean noise reduction was 14.2% at the level of the aorta and 14.5% at the level of the trachea on mediastinal images and 15.2% on lung images (ANOVA with Bonferroni correction; p < 0.0001); (b) leading to an overall noise reduction of 49.5% on lung images and a noise reduction of 45.7% at the level of the aorta and 51.7% at the level of the trachea on mediastinal images when compared with Group 1 images (ANOVA with Bonferroni correction; p < 0.0001).

Subjective image noise

The distributions of subjective image quality scores in Group 1, 2a and 2b are summarized in Table 1 . There was a significant reduction in the level of subjective image noise on lung and mediastinal images of Group 2a compared with Group 1 images (p < 0.003), further reinforced on Group 2b images (Group 2b vs Group 1; p < 0.0001) (Group 2b vs Group 2a; p = 0.0006) where image noise was rated as minimal in 78% of mediastinal images and 100% of lung images (McNemar chi-square test with Bonferroni correction).

Evaluation of the overall image quality

Table 1 summarizes the distributions of the overall image quality scores, significantly improved on Group 2a images compared with Group 1 images (p = 0.0081) and on Group 2b images compared with Group 2a images (p < 0.0001) (McNemar chi-square test with Bonferroni correction). Figure 1 illustrates the reduction in image noise obtained with iterative reconstructions in an obese patient.

Fig. 1
figure 1

CT angiographic examination in a 65 year-old obese woman (BMI: 41.4 kg/m2) referred for mediastinal mass (140 kV; 120 eff mAs; DLP: 263 mGy.cm), illustrating the reduction in objective image noise with 3 and 5 iterations. a: FBP-reconstructed mediastinal image at the level of the aortic arch. Objective noise measured at the level of the trachea: 37.6 HU. Subjective noise: score 2 b: Same anatomical level as that of (a), reconstructed with 3 iterations. Objective noise measured at the level of the trachea: 27HU. Subjective noise: score 2 c: Same anatomical level as that of (a), reconstructed with 5 iterations. Objective noise measured at the level of the trachea: 21.8HU. Subjective noise: score 2

Lesion conspicuity

A total of 96 lung images with at least one elementary lesion of mild lung infiltration were selected in Group 1 at the level of which we identified focal areas of ground glass attenuation on 49 images, ill-defined micronodules on 11 images, well-defined micronodules on 10 images, septal lines on 5 images, non septal lines on 29 images, lung cysts on 4 images and emphysematous lesions on 31 images. The individual CT features identified on Group 1 images were always identified on the corresponding Group 2a and Group 2b images. Table 2 summarizes the comparative analyses of lesion conspicuity on paired lung images, showing: (a) a similar conspicuity of well-defined micronodules, septal and nonseptal lines, and cysts in Groups 1, 2a and 2b; (b) a better conspicuity of ground glass opacities (Fig. 2), ill-defined micronodules (Fig. 3) and emphysematous lesions (Fig. 4) on Group 2a images compared with Group 1 images; (c) a better conspicuity of ground glass opacities and emphysematous lesions on Group 2b images compared with Group 2a images.

Table 2 Comparative analysis of lesion conspicuity on lung images showing features of mild lung infiltration
Fig. 2
figure 2

CT angiographic examination in a 55 year-old man (normal BMI: 19.8 kg/m2) referred for lung tumor follow-up (100 kV; 90 eff mAs; DLP: 108 mGy.cm). a: FBP-reconstructed lung image at the level of the aortic arch, showing ground glass opacities in both lungs. Objective noise measured at the level of the trachea: 42.1HU. Subjective noise: score 2 b: Same anatomical level as that of (a), reconstructed with 3 iterations. Objective noise measured at the level of the trachea: 26.3HU. Subjective noise: score 2. c: Same anatomical level as that of (a), reconstructed with 5 iterations. Objective noise measured at the level of the trachea: 19.3 HU. Subjective noise: score 1. Note the improved conspicuity of ground glass opacities from (a) to (c)

Fig. 3
figure 3

CT angiographic examination in a 30 year-old man (normal BMI: 23.55 kg/m2) referred for exacerbation of chronic obstructive pulmonary disease (120 kV; 90 eff mAs; DLP: 151 mGy.cm). a: FBP-reconstructed lung image at the level of the right upper lobe, showing ill-defined micronodules. Objective noise measured at the level of the trachea: 45HU. Subjective noise: score 2 b: Same anatomical level as that of (a), reconstructed with 3 iterations. Objective noise measured at the level of the trachea: 28.3HU. Subjective noise: score 2. c: Same anatomical level as that of (a), reconstructed with 5 iterations. Objective noise measured at the level of the trachea: 20.6 HU. Subjective noise: score 1. Note the improved conspicuity of ill-defined micronodules from (a) to (c)

Fig. 4
figure 4

CT angiographic examination in a 59 year-old man (normal BMI: 19.3 kg/m2) referred for suspicion of acute pulmonary embolism (100 kV; 90 eff mAs; DLP: 118 mGy.cm) a: FBP-reconstructed lung image at the level of the left lower lobes, showing disseminated centrilobular emphysema. Objective noise measurd at the level of the trachea: 32.1HU. Subjective noise: score 1 b: Same anatomical level as that of (a), reconstructed with 3 iterations. Objective noise measured at the level of the trachea: 19.8 HU. Subjective noise: score 1. c: Same anatomical level as that of (a), reconstructed with 5 iterations. Objective noise measured at the level of the trachea: 14.5 HU. Subjective noise: score 1. Note the improved conspicuity of emphysematous lesions from (a) to (c)

Discussion

To our knowledge, this is the first study evaluating the performance of an iterative reconstruction technique in image space (IRIS) for chest imaging. Comparing FBP and iterative reconstructions generated from the same data sets, our results demonstrate that the iterative reconstruction technique significantly reduces the level of objective image noise on standard CT examinations of the chest obtained in a non selected population of adult patients. While 3 iterations reduced the image noise by 31.5% and 37.2% on mediastinal images and by 34.3% on lung images, the greatest noise suppression was found to be obtained with 5 iterations, enabling a noise reduction ranging from 45.7% to 51.7% on mediastinal images and 49.5% on lung images compared with the FBP technique. Analysis of the distribution of subjective image noise scores showed significantly better ratings for Group 2a reconstructions compared with Group 1 reconstructions, both for lung and mediastinal images. However, the highest scores were observed in Group 2b with a minimal image noise on all lung images and in 78% of mediastinal images. The moderate image noise, always observed on our reference images, can be explained by a systematic weight-adapted selection of the kilovoltage in our daily routine. It should be pointed out that it did not alter the diagnostic value of images.

Owing to the recent introduction of other iterative reconstruction techniques, a single study reported a clinical experience of a different iterative reconstruction technique (adaptive statistical iterative reconstruction –ASIR) in chest CT in which the authors retrospectively selected 2 groups of chest CT examinations, comparing 98 examinations reconstructed with the FBP with 54 examinations reconstructed using the ASIR [17]. Several methodological differences in the CT protocols and study designs can explain why the objective noise measured on the images reconstructed with 3 and 5 iterations in the present study is higher than that reported in the study of Prakash et al using the ASIR technique. One explanation is that the regions-of-interest for noise measurements on the descending aorta in both studies were positioned at anatomical levels with major differences in terms of x-ray absorption, Prakash et al choosing the level of the carina while our study chose the level of the ventricular cavities. More important to consider are the major differences in the CT protocols chosen in both studies. Whereas our examinations were obtained at 100 or 120 kVp and 90 eff mAs, the CT protocol in the patients’ weight category comparable to that of our population (i.e., the 61-90 kg category), included 120 kVp and a milliamperage ranging from 75 to 440. As a result, the mean radiation dose in their study was 10 mSv whereas our average effective dose was only 2.13 mSv, i.e., a 5-fold difference in routine clinical practice.

Our study included a specific evaluation of lesion conspicuity in our study, choosing the CT features of mild lung infiltration to test this qualitative aspect of image quality. While each elementary lesion was identified on the 3 series of images, the visual depiction of ground glass opacities, ill-defined micronodules and emphysematous lesions was improved on Group 2a compared with reference images. Further comparison between the two levels of iterative reconstructions showed greater conspicuity of ground glass opacities and emphysematous lesions on Group 2b images, confirming that reduction of the objective image noise reinforces the detectability of subtle abnormalities, as also recently reported by Prakash et al with the ASIR technique [18]. Although depiction of ill-defined micronodules can also be influenced by the overall image quality, their conspicuity was not found to be increased on Group 2b images, probably because of the limited number of images showing this pattern. This preliminary experience does not confirm a practical concern recently raised by Xu et al [19]. Considering that radiologists are used to reading images reconstructed with FBP algorithm, these authors suggested that statistical reconstruction might give an impression of somewhat reduced diagnostic value.

This study suffers from several limitations. First, the evaluation of the iterative reconstruction technique was limited to a single CT protocol, based on single energy acquisitions. This constraint was due to the fact that, at the time of initiation of this evaluation, the prototype workstation was only able to reconstruct single energy data sets. A second limitation is that we restricted our investigation to 3 and 5 iterations and did not evaluate a wider range of iterations. This resulted from vendor-recommendations to avoid potential degradations of image texture and loss of spatial resolution when using more than 5 iterations. Third, the evaluation of lesion conspicuity was limited to the comparative analysis of CT features of lung infiltration and did not include mediastinal structures. However, the categories of lung lesions selected for this evaluation are among the most difficult to depict and one can reasonably assume that they reflect the detectability of subtle lesions on chest CT examinations reconstructed with IRIS. Lastly, image analysis was made by consensus between two readers and did not include the assessment of inter- nor intraobserver agreement between the two radiologists enrolled in this analysis. Owing to the recent introduction of iterative reconstructions, this study design was considered most suited for a preliminary evaluation.

In conclusion, our study shows that, compared with regular FBP, iterative reconstructions enable significant reduction of image noise without loss of diagnostic information, thus having the potential to decrease radiation dose during chest CT examinations.