Abstract
Objective
To compare fixed-flexion radiography of the knee with fluoroscopic semi-flexed radiography in terms of the reproducibility of measurements of minimum joint-space width (JSW) in the medial femorotibial joint.
Methods
Posteroanterior radiographs of the right knees of 18 normal volunteers were acquired with the patients standing on an upright fluoroscopy table, the feet externally rotated 10° and the toes touching the vertical table. Knees were positioned and radiographed with two different techniques: (1) semi-flexed positioning under fluoroscopic guidance using a horizontal X-ray beam; and (2) fixed-flexion positioning, with the knees and thighs touching the vertical table, using 10° caudal beam angulation without fluoroscopy. Foot maps were drawn in each case. Subjects were repositioned and radiographed twice using each technique. The posteroanterior beam angle that optimally projected the medial tibia plateau with the patient in the fixed-flexion position was also determined for each subject in a separate examination using fluoroscopy. Ten patients with osteoarthritis were also examined with the fixed-flexion technique using a conventional radiographic unit. Minimum medial joint-space width (JSW) in the medial femorotibial joint was measured manually with a graduated lens and also with a semi-automated computer algorithm.
Results
Reproducibility errors (root-mean-square SD) for manual and automated JSW measurement were 0.2 mm and 0.1 mm, respectively, for fluoroscopic semi-flexed positioning in volunteers; 0.3 mm and 0.1 mm, respectively, for fixed-flexion positioning in volunteers; and 0.2 mm and 0.1 mm, respectively, for fixed-flexion positioning in osteoarthritic patients. The optimal beam angle for visualizing the joint space was 9.0°±3.6°.
Conclusion
Fixed-flexion, non-fluoroscopic radiography of the knee can provide reproducible JSW measurement using widely available X-ray equipment. This technique is more feasible for multicenter clinical studies and routine clinical use than are methods that rely on fluoroscopic alignment of the tibial plateau.
Similar content being viewed by others
References
Peterfy C. Imaging techniques. In: Klippel J, Dieppe P, eds. Rheumatology, 2nd edn, vol 1. Philadelphia: Mosby, 1997:14.1–14.18.
Buckland-Wright JC, Macfarlane DG, Jasani MK, Lynch JA. Quantitative microfocal radiographic assessment of osteoarthritis of the knee from weight bearing tunnel and semiflexed standing views. J Rheum 1994; 21:1734–1741.
Mazzuca S, Brandt K, Katz B. Is conventional radiography suitable for evaluation of a disease-modifying drug in patients with knee osteoarthritis? Osteoarthritis Cart 1997; 5:217–226.
Mazzuca S, Brandt K, Buckland-Wright J, et al. Field test of the reproducibility of automated measurements of medial tibiofemoral joint space width derived from standardized knee radiographs. J Rheumatol 1999; 26:1359–1365.
Duryea J, Li J, Peterfy C, Gordon C, Genant H. Trainable rule-based algorithm for the measurement of joint space width in digital radiographic images of the knee. Med Phys 2000; 27:580–591.
Mazzuca S, Brandt K, Buckwalter K. Knee pain reduces joint space width in conventional standing anteroposterior radiographs of osteoarthritic knees. Arthritis Rheum (in press).
Buckland-Wright JC, Clarke GS, Chikanza IC, Grahame R. Quantitative microfocal radiography detects changes in erosion area in patients with early rheumatoid arthritis treated with myocrisine. J Rheumatol 1993; 20:243–247.
Buckland-Wright J, Wolfe F, Ward R, Flowers N, Hayne C. Substantial superiority of semiflexed (MTP) views in knee osteoarthritis: a comparative radiographic study, without fluoroscopy, of standing extended, semiflexed (MTP), and schuss views. J Rheumatol 1999; 26:2664–2674.
Acknowledgement
Portions of this work were funded by a grant (R01 AR43791-03) from the National Institute of Arthritis and Musculoskeletal and Skin Diseases (NIAMS).
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Peterfy, C., Li, J., Zaim, S. et al. Comparison of fixed-flexion positioning with fluoroscopic semi-flexed positioning for quantifying radiographic joint-space width in the knee: test-retest reproducibility. Skeletal Radiol 32, 128–132 (2003). https://doi.org/10.1007/s00256-002-0603-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00256-002-0603-z