Table 4

Evaluation of performance when comparing automatic diagnosis of AcuPebble SA100, based on ODI alone considering ≥3% desaturations as the threshold for events, with the reference test (CR-PG)

StatisticValue95% CI
Disease prevalence (%)52.0043.70% to 60.22
Sensitivity (%)91.0382.38% to 96.32
Specificity (%)93.0684.53% to 97.71
Positive likelihood ratio13.115.61 to 30.62
Negative likelihood ratio0.100.05 to 0.20
Positive predictive value (%)93.4285.87 to 97.07
Negative predictive value (%)90.5482.48 to 95.11
Accuracy (%)9286.44 to 95.8
Cohen’s Kappa0.840.75 to 0.93
  • CR-PG, cardiorespiratory polygraphy; ODI, Oxygen Desaturation Index.