Table 1

MRI acquisition parameters

DWIT2-FLAIRT2* GRE 2DSWI 3D (see note 1)3D T1-volumeT22D T1 (see note 2)
Orientation AxialAxialAxialUse standard manufacturer specific SWI protocol from manufacturer’s protocol tree (ie, SWAN (General Electric); VEN_BOLD (Philips); SWI (Siemens))SagittalAxialAxial
Plane 2D—EPI2D2D3D2D2D
TE (ms) Minimum125–14020–30
(15–40)
Minimum85–10010–14
TR (ms) Minimum11 000300–1000Minimum3000–5600600–650
TI (ms) NA2800NA450– 1000
(or no inversion on old General Electric FSPGR)
NANA
Slice thickness (mm) ≤4.0≤4.03
(3-5)
1 mm isotropic voxels
(1.25 mm)
≤4.0≤4.0
Slice gap (mm) ≤0.4≤0.4≤0.3
(0–1)
NA≤0.4≤0.4
Acquisition matrix (RFOV acceptable) 96–128180–256
(180-512)
180–256128–256180–256
(180-512)
180–256
FOV (mm) 230–240230–240230–240224–256230–240230–240
Flip angle (excitation) 90°90°15–60°8–15°90°90°
Flip angle (refocusing) 180°180°NANA180°180°
b-value 1000NANANANANA
Number of slices 32–4232–4232–42180–192
(128–256)
32–4232–42
  • Preferred parameters are listed, with acceptable range of values given in brackets for sites where the preferred parameters cannot be achieved for technical reasons .

  • Note 1: If available.

  • Note 2: Only required if 3D T1-volume acquisition is degraded by patient motion.

  • DWI, diffusion-weighted image; EPI, echo-planar imaging; FLAIR, fluid-attenuated inversion recovery; FOV, field of view; FSPGR, fast spoiled gradient-echo; GRE 2D, gradient echo two dimensional; NA, not applicable; RFOV, rectangular field of view; SWI 3D, susceptibility weighted imaging three dimensional; TE, echo time; TI, inversion time; TR, repetition time.