Table 1

Sensitivity, specificity, PPV, NPV and κ for automated urinalysis versus the laboratory reference test (Urisys 2400), as well as for visual urinalysis versus the laboratory reference method

Urisys 2400Sensitivity (95% CI)Specificity (95% CI)PPV (95% CI)NPV (95% CI)κ (0.95 CI)
Visual
 NitritePositiveNegative
 N=234Positive441496% (85% to 99%)93% (88% to 96%)76% (63% to 86%)99% (96% to 100%)0.803 (0.710 to 0.896)
Negative2174
Automated
 Nitrite
 N=227Positive40991% (78% to 97%)95% (91% to 98%)82% (68% to 91%)98% (94% to 99%)0.824 (0.732 to 0.917)
Negative4174
Visual
 LeucocytesPositiveNegative
 N=234Positive125561% (54% to 67%)82% (63% to 94%)96% (91% to 99%)22% (15% to 31%)0.197 (0.062 to 0.332)
Negative8123
Automated
 Leucocytes
 N=227Positive130465% (58% to 72%)86% (67% to 96%)97% (93% to 99%)26% (17% to 36%)0.256 (0.115 to 0.396)
Negative6924
Visual
 ErythrocytesPositiveNegative
 N=230Positive79856% (47% to 64%)91% (83% to 96%)91% (83% to 96%)56% (47% to 64%)0.416 (0.303 to 0.529)
Negative6380
Automated
 Erythrocytes
 N=229Positive91865% (56% to 73%)91% (83% to 96%)92% (85% to 96%)62% (53% to 71%)0.517 (0.408 to 0.626)
Negative4981
  • NPV, negative predictive value; PPV, positive predictive value.