Modelling and in vivo evaluation of tendon forces and strain in dynamic rehabilitation exercises: a scoping review
  1. Adrian Escriche-Escuder1,2,
  2. Antonio I Cuesta-Vargas1,2,3,
  3. Jose Casaña4
  1. 1Department of Physiotherapy, University of Malaga, Malaga, Spain
  2. 2Instituto de Investigación Biomédica de Málaga (IBIMA), Malaga, Spain
  3. 3Department of Health, Queensland University of Technology, Brisbane, Queensland, Australia
  4. 4Department of Physiotherapy, University of Valencia, Valencia, Spain
  1. Correspondence to Dr Antonio I Cuesta-Vargas; acuesta{at}uma.es

Abstract

Objectives Although exercise is considered the preferred approach for tendinopathies, the actual load that acts on the tendon in loading programmes is usually unknown. The objective of this study was to review the techniques that have been applied in vivo to estimate the forces and strain that act on the human tendon in dynamic exercises used during rehabilitation.

Design Scoping review.

Data sources Embase, PubMed, Web of Science and Google Scholar were searched from database inception to February 2021.

Eligibility criteria Cross-sectional studies available in English or Spanish language were included if they focused on evaluating the forces or strain of human tendons in vivo during dynamic exercises. Studies were excluded if they did not evaluate tendon forces or strain; if they evaluated running, walking, jumping, landing or no dynamic exercise at all; and if they were conference proceedings or book chapters.

Data extraction and synthesis Data extracted included year of publication, study setting, study population characteristics, technique used and exercises evaluated. The studies were grouped by the types of techniques and the tendon location.

Results Twenty-one studies were included. Fourteen studies used an indirect methodology based on inverse dynamics, nine of them in the Achilles and five in the patellar tendon. Six studies implemented force transducers for measuring tendon forces in open carpal tunnel release surgery patients. One study applied an optic fibre technique to detect forces in the patellar tendon. Four studies measured strain using ultrasound-based techniques.

Conclusions There is a predominant use of inverse dynamics, but force transducers, optic fibre and estimations from strain data are also used. Although these tools may be used to make general estimates of tendon forces and strains, the invasiveness of some methods and the loss of immediacy of others make it difficult to provide immediate feedback to the individuals.

  • Musculoskeletal disorders
  • SPORTS MEDICINE
  • REHABILITATION MEDICINE
  • Foot & ankle
  • Hand & wrist

Data availability statement

No data are available. No additional data available.

http://creativecommons.org/licenses/by-nc/4.0/

This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/.

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Strengths and limitations of this study

  • The extensive search carried out in this review in four of the main databases allows the reader to approach a wide field of knowledge.

  • This review provides a summary of the available literature on the study of forces and strain that act on the tendon during dynamic exercises.

  • Grouping the assessment tools into subgroups allows an analysis of the advantages and disadvantages of each option.

  • Some studies might not have been identified due to the difficulty in tracking the literature because of the variety of terms used.

Introduction

Tendinopathy is the preferred term for persistent tendon pain and loss of function related to mechanical loading.1 The high incidence and prevalence of this disorder alters the ability of people to work, exercise or perform activities of daily life, causing a great social and economic burden.2

Current knowledge supports the need to integrate an active approach for tendinopathy, based on a conservative management that includes education, exercise (with appropriate management and modification of loads) and support interventions for pain and symptom control.2 Thus, loading interventions with a progressive exercise programme are considered an essential part of the management of tendinopathies due to the vast evidence published in the last decades.2–6 These approaches focus on producing an adequate stimulus for tendon adaptations and aim to increase the patients’ loading capacity.3 7 Regarding the adaptations in the tendon, research data suggest that tenocytes respond to mechanical loading by inducing anabolic and catabolic processes of matrix proteins, respectively, through a process known as mechanotransduction.7–11 Therefore, tendon strain is an important factor for the maintenance and adaptation of the tissue.

Different exercise modalities and intensities have been applied in tendinopathy with reasonably good results.6 12–14 Likewise, different strategies have been implemented for handling and modifying the applied loads.15–17 However, although some concepts such as repetition maximum have made it possible to parameterise and quantify the applied dose based on the subject’s ability to perform an activity a specific number of repetitions, the actual load that acts on the tendon in these activities is usually unknown. In both prevention and treatment of tendinopathy, load management would benefit from a greater understanding of the loads that act on the tendon during exercises and the strain that occur under load, especially considering that there may be a ‘sweet spot’ of tendon strain for stimulating adaptation.7

In the analysis of the loads that act on the tendon, it is relevant to differentiate between physical quantities such as force and strain. Tendon force is a measure of the absolute load that acts on the tendon, while strain refers to the deformation of the tendon relative to its resting state. Strain has a different nature depending on the force that produces it. Thus, tendons are subjected to compression, tension or shear forces in daily activities,18 19 but it is the tensile load (and the strain it produces) that plays a leading role in the function of the tendon.20 Therefore, the evaluation of the tensile strain is especially relevant for the study of the loading programmes.21

Regardless of the parameter evaluated, it is important to take into account a factor that makes studying in vivo tendon mechanics difficult: tendons are not uniaxial structures but are usually made up of different bundles.22 This causes regional variations in mechanical properties, and the distribution of forces and strains throughout the tissue is not uniform.23 Tendon forces have been calculated through in vitro studies,24 as well as have been estimated through in vivo indirect calculations based on body position, joint reaction forces and inverse dynamic models.25–27 Additionally, as underlined by a previous review, invasive evaluations using force transducers and optic fibre techniques have enabled the direct measurement of forces in tendons of the hand and the Achilles and patellar tendons.23

Medical imaging techniques such as ultrasound or MRI have previously made it possible to directly measure strain during isometric contractions,28 walking,29 30 running27 31 and hopping.32 However, transducer position may affect the ultrasound measurements significantly, and it is necessary to use a rigid fixation over the tissue that may alter movement patterns.33 Therefore, its use in some dynamic activities is still limited.

Some reviews have been previously published focused on the evaluation of tendon loads.23 33 These reviews are not specific to dynamic rehabilitation exercises and include mainly methods developed for the study of isometric contractions28 34 or cyclic activities such as running,35 36 cycling25 37 or walking.38 39 Some of these methods have been adapted to the study of dynamic exercises (such as rehabilitation exercises), but the study of this type of exercises is still scarce due to the limitations of these tools.33 Therefore, there is still a lack of studies addressing the direct measurement of loads and the evaluation of dynamic exercises commonly used during rehabilitation processes.

The aim of this study is to review the techniques that have been applied in vivo to, directly and indirectly, estimate the forces and strain that act on the human tendon in dynamic exercises commonly used during rehabilitation processes.

Methods

This scoping review was undertaken following the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) Extension for Scoping Reviews (PRISMA-ScR) guidelines.40 This review has not been registered in PROSPERO because this platform does not currently accept registrations for scoping reviews, literature reviews or mapping reviews.

Information sources and search strategy

According to the recommendations of a recent study41 for biomedical reviews, four databases were searched by two reviewers (AE-E and JCG) from database inception to February 2021: Embase, PubMed (including Medline), Web of Science and Google Scholar. The following combinations of terms were used in the first three databases: “Tendon (Title) AND Load (Title)”; “Tendon (Title) AND Force (Title)”; “Tendon (Title) AND Biomechanics (Title)”; “Tendon AND wave”; “Tendon (Title) AND Properties (Title)”. Additionally, “Tendon AND Load” was searched in Embase and PubMed. The combinations of terms “Tendon AND Force”, “Tendon AND Biomechanics”, “Tendon AND Properties”, “Tendon AND Load” and “Tendon AND wave” were used in Google Scholar, retrieving the first 200 relevant references of each search. Detailed information on the sources of information and the combinations of terms used is available in the online supplemental appendix 1.

Eligibility criteria

All studies that met the following eligibility criteria were included:

  1. Cross-sectional studies published in scientific journals.

  2. Focused on evaluating the forces and strain (tendon strain evaluation was included if it was described as a way to quantify loads) of tendons in vivo using direct or indirect techniques.

  3. During dynamic exercises.

  4. Available in English or Spanish language.

Conversely, those studies meeting any of these exclusion criteria were discarded: (A) studies with evaluation of neuromuscular or joint forces that do not describe evaluating the tendon; (B) investigated tasks were running, walking, jumping, landing or other everyday tasks that are not rehabilitative exercises; (C) conference proceedings; and (D) book chapters.

Study selection

All retrieved references were imported into Mendeley to later be included in Rayyan (https://www.rayyan.ai/), a systematic review support tool. Duplicates were identified and removed. The remaining references were screened by title and abstract by one author (AE-E) to exclude clearly irrelevant articles. Finally, two reviewers (AE-E and JC) screened the full texts of identified articles to select those that met the eligibility criteria. A third reviewer solved any disagreements (AIC-V).

Data extraction

Two reviewers (AE-E and JC) assessed the full texts of the selected studies. To obtain the information from the studies, an extraction form was used including the following data: authors and year of publication; study setting; study population; participant demographics; details of the evaluation technique; dynamic exercises evaluated; and tendon forces/strain results.

In this review, they were included those studies that analysed the forces and strain on the tendon in dynamic exercises, especially those commonly used in tendon rehabilitation. Dynamic analysis based on running, walking or cycling, and batteries of exercises based on day-to-day or work activities were not taken into account.

Synthesis of results

The studies were grouped by the types of measurement techniques applied and by the tendon location, summarising the type of settings, populations and article types for each group, along with the broad findings.

Methodological quality

Current guidelines on conducting a scoping review describe the inclusion of a methodological quality analysis as not necessary.42 43 Likewise, the lack of a standardised tool for the methodological evaluation of the heterogeneous type of studies included in this review makes methodological analysis difficult. In this context, this review focus on analysing the forces and strain evaluation methodologies used in the included studies rather than in the magnitude of the results obtained, with the lack of methodological quality analysis influencing the results and conclusions of this review to a lesser extent.

Patient and public involvement

None.

Results

A total of 16 571 records were identified in PubMed, Embase, Web of Science and Google Scholar. Then, duplicates were removed, remaining 8536 references. Additionally, eight records were identified by additional sources. Among these, 153 were identified as potentially eligible after reading the title and the abstract, retrieving the full texts of all of them. After evaluating the fulfilment of the eligibility criteria, 21 studies were finally included in the current review. The figure 1 represents the flow diagram of the selection process. A detailed list of the studies excluded in the last stage is available in the online supplemental appendix 2.

Figure 1

Flow diagram of the selection process.

In total, 300 subjects were included in the analysed studies. Among these, 202 correspond to healthy samples, while 98 of them were open carpal tunnel release surgery patients. However, due to the similarity in the characteristics of the sample and the concurrence of most of the authors in the case of three studies44–46 (12 subjects in each study), it is pertinent to think that they are the same participants.

Modelling and in vivo evaluation methodologies

Different evaluation methodologies were identified in the included reports, including inverse dynamics, force transducers and optic fibre sensors for the evaluation of tendon forces, and ultrasound imaging techniques for strain evaluation. The tendon locations evaluated were the Achilles, quadriceps, patellar and different tendons of the hand. Table 1 shows the groups of evaluation techniques associated with the tendon location and the references of the records that included each one. Table 2 includes expanded information about the measurement methodology.

Table 1

Forces and strain evaluation methodologies identified in the included studies

Table 2

Characteristics of the included studies

Force

Inverse dynamics

Fourteen studies used an indirect evaluation methodology of tendon forces based on inverse dynamics, nine of them in the Achilles tendon and five in the patellar tendon. When inverse dynamics are used, tendon forces are estimated using different equations based on joint torque and moment arms or integrating kinematic and kinetic data in musculoskeletal models. This methodology uses kinematics, often complemented with applied external forces, to calculate net joint moments.47 Moment arms are estimated from previous literature data or estimated specifically for each patient through imaging techniques such as MRI or ultrasound.

Most of the included studies used motion capture systems for kinematics, while force plates were the most used device for obtaining kinetic data. Some studies used generic moment arms based on the published literature,48 49 other used previously described procedures and equations,50–53 while other estimated subject-specific moment arms based on imaging techniques.54 55 Kinematic and kinetic data were integrated into different musculoskeletal models: three studies56–58 used the Human Body Model,59 one48 study used the OpenSim model,60 one study54 used the FreeBody model,61 while other studies52 55 62 implemented other codes or models.

Most of the studies reported normalised force values by body weight (BW), obtaining the lowest values in the Achilles through the seated heel raising exercise (0.41–0.5 BW).48 54 The single-leg heel raising and lowering obtained values between 3–5.12 BW for the Achilles tendon.48 54 56 62 In the patellar tendon, the results were mainly reported in Newtons (N), obtaining mean values between 2899 and 5683 N for different variants of the squat.49 53

Force transducers

Six studies implemented force transducers for measuring tendon forces, all of them in open carpal tunnel release surgery patients. The introduction of the force transducers was carried out during surgery with local anaesthesia. Three modalities of force transducers were applied: buckle force transducer,44–46 S-shaped force transducer63 and load cell.64 65

The buckle force transducers technique used in three of the studies44–46 consisted of a modified version of the method described by Dennerlein et al.66 This device consisted of a 9×16 × 4.5 mm stainless steel frame and a removable fulcrum designed to fit inside the carpal canal.44–46 In this system, each tendon lies in semicircular arches in the device.44–46 These studies evaluated unresisted finger flexion and extension at different wrist angles, obtaining a range of mean values between 1.3 N and 25.5 N for the flexor digitorum profundus (range −1.6 N to 74.7 N) and 1.3 N – 12.9 N (range −2.0 N to 47.53 N) for the flexor digitorum superficialis.44–46 The S-shaped force transducer consisted of a stainless steel frame combined with four strain gauges attached on its central beam.63 This study obtained values between 0 and 12.0 kgf (117.7 N, obtained with the active tip pinch) in the evaluation of different finger and wrist flexion and extension exercises.63 In the case of load cell, an apparatus consisting of three vertical rods, each terminating in a ‘hook’ was used for the tendon force measurements.64 65 The central hook was connected to a load cell, recording the applied forces. These studies evaluated different finger flexion and extension exercises, with and without resistance, obtaining values in a range between 1 N and 50 N (resisted finger flexion, 300 g).64 65

Optic fibre sensor

Dillon et al67 applied an optic fibre technique to detect forces in both the anterior and the posterior regions of the proximal patellar tendon. This methodology was implemented inserting two 0.5 mm optic fibre sensor perpendicular through the entire cross section of the tendon under local anaesthesia. For the purpose of the study, one sensor was placed 1–2 mm anterior to the posterior border of the tendon, while the other sensor was placed 1–2 mm posterior to the anterior border of the tendon.67 The optic fibre was attached to a transmitter–receiver unit for light intensity monitoring. Then, tendon forces were registered during dynamic exercises, removing the sensor at the end of all tests.67 In this study, the sensors were not calibrated to record forces in N. Therefore, the data are only available through the differential output of the fibre signal.67 In general, higher values were found in the posterior area of the proximal tendon (0.77–1.00 V) than in the anterior area (0.21–0.42 V). The highest values were found in the one-legged squat exercise (1.00 V).67

Strain as a load measure

Four studies55 57 68 69 carried out additional measurements for quantifying loads on the tendon through strain or elongation measurement. Rees et al68 and Chaudhry et al55 calculated the Achilles tendon length as the distance between the medial gastrocnemius myotendinous junction (tendon origin) and the tendon insertion, using ultrasound imaging. Rees et al68 established and tracked the position of these anatomical sites in terms of three-dimensional (3D) coordinates over time by using an active marker motion analysis system through a previously detailed methodology.32 Chaudhry et al55 implemented an algorithm that provides an intensity map of the ultrasound images, from which the two-dimensional (2D) position and angular orientation of the most intense points can be established.55 Thus, the authors used this mechanism to locate and track the myotendinous junction.55 Elongation was calculated as the difference between the instantaneous length and the initial length. In these studies, standing eccentric heel-drop and concentric heel-raises exercises were assessed, both phases performed with bent and extended knee.55 68 In the study by Rees et al, the authors found that the elongation of the tendon during the eccentric and concentric part of the exercise is similar (13.6 mm and 14.9 mm on average for eccentric and concentric phase, respectively).68 Chaudhry et al55 also obtained similar elongation of the tendon during the eccentric (approximately 8 mm of peak mean elongation) and concentric phase (approximately 7 mm).55 Earp et al69 estimated the myotendinous unit length of the distal vastus lateralis using previous models based on joint position and individual limb lengths. This information was used to compare different ways of performing the squat, analysing the tendon lengthening pattern during the concentric and eccentric phases based on the muscle fascicle behaviour.69 Revak et al57 estimated the tendon strain using the average Young modulus value (819 N/mm2) reported in previous literature.70 First, the Achilles tendon stress (magnitude that quantifies the load per unit area of the tendon) was calculated by dividing the tendon force (estimated using inverse dynamics) by the cross-sectional area of each participant.57 Then, the tendon strain was calculated by dividing the tendon stress by the Young modulus. In this case, ultrasound was used to measure the cross-section of the tendon (not during exercises).57 The strain values obtained (expressed in %) were between 0.71±0.35 and 8.80±0.35, corresponding to the seated heel raising and lowering and the unilateral heel raising and lowering exercises, respectively.57

Type of exercises

Different types of exercises were analysed in the included studies. Heel raising and lowering exercises, involving concentric or eccentric plantarflexion, are commonly applied in Achilles tendinopathy rehabilitation. Seven studies included this type of exercises.48 54–57 62 68 In patellar tendon disorders, different modalities of squats are commonly prescribed, as well as exercises involving knee flexion and extension. Eight48–50 52 53 56 67 69 and two51 67 studies analysed these types of exercises, respectively. Another exercise commonly applied for lower limb disorders such as lunge was analysed in two studies.56 58 Three studies analysed step-up and step-down exercises or stairs climbing.48 51 67 Table 2 includes the type of exercises analysed in each study.

Discussion

The aim of this study was to review the techniques that have been applied in vivo to estimate the forces and strain that act on the human tendon in dynamic exercises commonly used during rehabilitation processes. The main finding of this review is that most studies used an indirect method such as inverse dynamics, while there is a lack of direct measurements due to the difficulties and limitations in its application.

Indirect force measurement: inverse dynamics

Most of the studies included in this review used inverse dynamics as an indirect evaluation of tendon forces. This methodology uses measured kinematics and external forces to indirectly calculate net joint torques and forces in a body segment model.71 These calculations are usually based on the joint moments produced by the muscle or muscles to which the tendon is inserted. Then, the biomechanical study is based on a single agonist force vector in line with the tendon direction and, in some cases, on a single antagonist force vector in the opposite direction.72 Although this method is widely used, it is suggested that the results obtained differ from the actual ones due to incorrect modelling assumptions and measurement errors.71 For example, classical inverse dynamics assumes idealised pin joints and the existence of rigid body segments and that does not match reality.71 Kinetics are introduced in the procedures with the intention of limiting these errors. However, due to the aforementioned difficulties of kinematics measurements, the kinematics and kinetics data are not always consistent. This creates a new problem due to the concurrency of data that does not match, forcing part of the data to be discarded.71

There are different procedures based on inverse dynamics for the calculation of forces. Thus, although most of the included studies used similar kinematics (motion capture devices) and kinetics (force plates) assessment systems, these data were processed in different ways. Some studies integrated these data in musculoskeletal models such as Human Body Model,56–58 OpenSim,48 FreeBody,54 among others.52 55 62 These models make more or less precise assumptions that allow us to transform the kinematics and kinetics data into net torques of body segments. Likewise, models such as the Human Body Model made an additional indirect estimate, first calculating the muscle forces and assuming that the forces in the tendon will be equal to the sum of the muscle forces of the agonist muscle group.56–58 This fact could imply an additional error in the estimation since there may be differences between the agonist muscle group and tendon forces, and a potential error is made when only some of the muscles involved in the movement are taken into account.72 Different methods were used for estimating the moment arms. Some musculoskeletal models used previous estimations of the moment arms, with some differences both in the models and in the equations used.52 55 62 Some studies performed subject-specific calculations based on imaging techniques to minimise error,54 55 and other studies used data from previously published literature (eg, 5 cm ankle moment arm).48 49 Alternatively, some studies used an intermediate method based on the use of new or previously published equations together with specific data from each patient.50–53 Thus, the results obtained may be influenced by the specific limitations of each methodology. Using generic moment arms based on normative data ignores anatomical differences between individuals,72 73 and, sometimes, this value is not scaled to the rest of the anatomical structures.72 74 Previous studies also suggest that the moment arm cannot be estimated from easily measured anthropometric characteristics or joint size differences, supporting the use of imaging techniques.75 In cases where the moment arm is directly measured, it should be noted that the values in a resting position may not correspond to the values in another position or to those that would be obtained with the addition of muscle contraction.72 74 The chosen method is relevant because, according to previous studies, there could be differences of up to 40%–50% depending on the technique used (for the patellar tendon moment arm length at a knee angle of 90°).72 73 Likewise, these differences could translate into up to 67% differences in the estimated values of tendon force.72 73

Despite all the previously mentioned limitations, modelling approaches have been widely employed to estimate tendon forces.47 This may be due to its main advantage: it is a non-invasive procedure.

Direct force measurement

In the last decades, an attempt has been made to develop direct measurement techniques. However, this approach is limited due to the need to insert sensors into the body. This characteristic makes it a highly invasive procedure, making its use in healthy subjects difficult to justify.33 Sensors must be biotolerable (for short-term measurements) and biocompatible (for long-term use), as well as easy to implant.23 Additionally, devices should avoid damaging body tissues and alter the tendon and joint mobility and neuromuscular function.23 It has been suggested that these sensors should also be flexible and allow wireless data transmission to facilitate their clinical use.33 The transducers are implanted with an incision of several centimetres. Thus, the wound usually impedes normal activity for 2–3 weeks and sometimes makes it difficult to measure activity during the same session in which the sensor is inserted.76 Additionally, potential complications such as local pain or infections have limited the use of this methodology to a restricted research population.76

Force transducers

Buckle transducers were one of the first devices to show a successful ability to directly assess these forces in various activities such as walking, running, cycling or jumping.37 77–79 This kind of transducer consists of a metallic buckle with strain gauges through which a tendon is looped.23 When a tensile force is applied to the tendon, the buckle deforms and produces a voltage output proportional to the force.23 Due to their configuration, these buckle transducers enable the measurement of force of the entire cross-section of the tendon.23 This is an advantage over other implantable transducers (eg, optic fibre) that only record forces in a specific area, since it is known that the load may not be uniformly transmitted throughout the entire tendon section.23 80–82 However, the placement of the tendon through the buckle shortens the tendon and can alter its natural movement.23 Additionally, small changes in the placement may cause measurement differences, so it is recommended to carry out the calibration of these transducers within the specific tissue under study, and once the sensor is placed and calibrated, it should be avoided to modify or remove it until the measurement is finished.23

In this review, six studies introduced force transducers for measuring tendon forces during wrist and fingers flexion and extension rehabilitation exercises, all of them in open carpal tunnel release surgery patients. Taking advantage of surgery to place the sensor makes it possible to compensate for part of the invasiveness that this procedure entails. However, reducing its application to this context limits the contexts in which it may be applied. In this regard, the development of biodegradable sensors that are reabsorbed after a certain time could increase the situations where their application can be justified, since the avoidance of a second surgery to remove the sensor would reduce some drawbacks of the technique.83 In all cases, the procedure was carried out after the application of anaesthesia, which together with the surgical procedure itself could have some impact on the measurement results.

Optic fibre sensor

The use of optic fibre sensors appeared as a smaller solution compared with previous force transducers.84 This kind of sensor is inserted perpendicular through the tendon. When a longitudinal tension is produced in the tendon, negative transverse tension is produced that squeezes the optical fibre.23 85 The functioning of the optical fibre sensor is based on the amplitude modulation of the transmitted light that occurs when the optical fibre changes its shape due to the forces acting on it.23 85 These differences can be seen in the receiver, which provides a voltage output proportional to the intensity of the light detected and therefore related to the tendon tensile strain.23 85 This effect can be achieved using two types of sensors: intensity-based and spectral-based optical sensors.76

During the last decades, different devices based on optic fibre have been developed and applied to directly measure tendon forces in vivo in humans during isometric contractions86 and during dynamic activities such as walking or jumping.39 76 84 87 88 These sensors have evolved from the earliest models (approximately 500 µm)89 to modern spectral-based models incorporating fibre Bragg gratings and microfabricated stainless steel housings (approximately 200 µm).76 Modern optic fibre sensors offer some advantages such as small size, high sensitivity, fast response time, large dynamic range and insensitivity to electromagnetic interference.76 However, the main limitation of this measurement technique is still the invasiveness of the procedure for introducing and removing the sensor.76 The procedure is usually performed under local anaesthesia, causing a little wound in the tissue that can interfere with movement.76 Due to its smaller size, compared with the buckle transducer, the insertion process, the wound and the recovery process are of lesser magnitude. Thus, its use in volunteers is more easily justified.23 Also, the possible interference of the sensor during movement and changes in the natural shape of the tendon are reduced compared with other transducers, although still existing.23 84

This technique has other limitations to take into account. Previous studies have found that skin movement, cable migration and loading rate may influence the accuracy of the sensor.89 Therefore, this technology may be considered an appropriate option for in vivo evaluation as long as these artefacts can be minimised.76

Furthermore, this kind of sensor records forces in a specific area of the tendon, and this could be a source of differences between measurements due to the fact that force may not be uniformly transmitted throughout the entire tendon section.23 80–82 This phenomenon could be related to the relative sliding between the different tendon fascicles.80 81

The lack of studies using this technique in dynamic exercise could be because of the current limitations that, although lower than those of other invasive techniques, still represent a significant barrier to its implementation. Thus, further study of the matter is encouraged.

Strain

Tenocytes are sensitive to strain.7 21 90 91 Thus, it has been suggested that it is the strain magnitude experienced by tendon fibres, not force, that is more directly related to the positive or negative effects triggered in the tissue.7 21 90 Previous studies have shown that tendon strain during activities such as walking or running is between 4.0%–4.3% and 4.6%–9.0%, respectively. The only study that reported the percentage of tendon strain in this review found a strain between 0.71% (seated heel raising and lowering) and 8.80% (standing unilateral heel raising and lowering exercises).57

The use of imaging techniques (eg, 2D28 92 93 and 3D94 ultrasound or MRI92) has been previously reported, especially during isometric contractions, but most of these methods have not been transferred to the study of dynamic rehabilitation exercises.

Tendon are viscoelastic, and their mechanical and viscoelastic properties of the tendon may imply a time-dependent behaviour of the tendon when a force is applied to it.11 69 However, the hysteresis of tendons has been reported to be approximately 10%,95 and the loading rate effect does not seem to be decisive in the range of loading rates applied during physical activities.96 97 Furthermore, current strain evaluation techniques (ultrasound-based methods) seem not to be sensitive enough to detect the small effects that this range of loading rates produces.98 To further minimise these loading rate effects, the application of conditioning contractions may allow a state of certain stability and reliability to be reached at the moment of the application of forces for its evaluation.99 100 However, this is not done or at least described in most studies.

In this review, four studies55 57 68 69 included a tendon elongation measurement for assessing tendon loads. Revak et al57 calculated the tendon strain by dividing the tendon stress (previously obtained) by the average Young modulus reported (819 N/mm2).57 This methodology again requires making various assumptions to estimate the tendon strain through the tendon stress, which in turn has been calculated using the tendon force value calculated indirectly using inverse dynamics. Therefore, this indirect method could accumulate the error of all the intermediate steps, some of which have been discussed in previous sections. Additionally, it also does not seem justified to assume a constant Young modulus for different individuals. Earp et al69 estimated the myotendinous unit length of the distal vastus lateralis using derived models based on joint position and individual limb lengths and calculated tendon lengthening based on muscle fascicle behaviour, a method that has been found to be reliable.101 102 Differently, Rees et al68 and Chaudhry et al55 calculated the Achilles tendon length as the distance between the tendon origin and the tendon insertion.68 Thus, they tracked the position of these anatomical sites by using an active marker motion analysis system.68 To do this, it is necessary to define what is the position for the initial length, also known as zero-length. Although the neutral position of the joint is often used as zero-length position,55 it should be noted that this position of the ankle seems to be already associated with longitudinal tendon strain, and the zero-length has been previously related to a different position (knee angle of 180° and ankle angle of 110°).103 Thus, the joint position corresponding to the zero-length is not always precisely known.23 It is important to normalise this parameter to allow comparison between studies, for example, using a standardised position of the joint.23 In these cases, we usually speak of ‘relative strain’ with respect to that previously determined position. While this methodology may be useful when comparing the peak strains of a tendon under different exercises within a particular study or with studies that use that same position, this methodology does not allow for comparing these results with those of in vitro studies, where the position of zero-length is precisely determined.23 The use of a force sensor in conjunction with ultrasonography could help determine the zero-length in each subject.23 Other limitations of the approach used in these studies are the skin movements and the curved path of the tendon. Previous evidence have found that considering the Achilles tendon as a straight line between gastrocnemius medialis myotendinous junction and calcaneus results in an underestimation of the tendon length and carries errors of up to 78% of the length changes.104 In this regard, Kharazi et al105 developed a new approach for Achilles strain in vivo measurement, which considers the tendon curve-path shape using skin reflective markers.

Imaging techniques

Ultrasonography as a strain measurement technique has some important advantages over other methods: it is non-invasive, does not expose the volunteers to radiation and it is relatively affordable.33 The absence of a sensor inside the body that can hinder mobility, together with the non-use of anaesthesia, allow natural movements.23 Additionally, ultrasonography enables the differentiation of muscle and tendon interfaces, enabling muscle and tendon strains to be independently measured.33 Basically, two approaches could be used to analyse strain using imaging techniques: on the one hand, displacement measurements between the tendon origin and insertion anatomical sites (myotendinous junction), approach used in this review by Rees et al68 and Chaudhry et al.55 The tracking of these anatomical sites is done through different methods. Initially, this task was performed through manual marking of the anatomical sites in successive ultrasound frames throughout the movement.55 However, this methodology was excessively laborious, so it was limited to only a few frames.55 For this reason, different algorithms, usually based on cross-correlation, have been developed to automate the process.55 106–108 In the Achilles tendon, for example, insertion is usually tracked using a marker placed on the calcaneus, while for the myotendinous junction, active marker motion analysis and ultrasound systems have been combined.55 68 On the other hand, displacement measurements between known points within the tendon midsubstance, known as speckle-tracking, can be used.33 The speckle-tracking technique allows unique speckle patterns of the tendon to be identified and tracked during movement.109 The regional strain measurement approach is an advantage over implantable sensors that only enable point-to-point strain assessment. The choice of approach is important since, taking into account that the strain distribution is not consistent throughout the tendon, the result may also be different. While the first option provides the value of the global strain across the entire length of the tendon, the second one offers a measure of a specific region. Some studies have reported that the displacement of the proximal insertion point may be a representative measurement of the total tendon elongation during contraction, but more recent works have shown the limitations of this approach.72 Thus, both methods may be adequate as long as they are properly reported, only being possible to compare results from the same approach.23 Likewise, the choice of the anatomical site used as a tracking landmark is relevant. Thus, previous studies have shown that small variations (eg, tibial tuberosity or plateau) result in significant differences in the values obtained, both in tendon strain itself and in other calculated mechanical properties (eg, tendon stiffness).110 Numerous limitations of imaging techniques have been widely reported.23 33 72 It is worth emphasising that most of these limitations are already present in measurements during isometric contractions, making progress to the measurement of dynamic exercises even more challenging. First, the ultrasound probe placement and orientation may affect the measurements, and any motion produced during the body segment movement can be a source of error.33 72 111 In the case of the study of isometric contractions, researchers have tried to overcome this limitation by means of rigid fixation with straps. However, this fixation is difficult to achieve during dynamic exercises and, especially during great joint angle excursions, it is difficult to maintain a stable image of the tendon or myotendinous junction. Additionally, the fixation can interfere with the movement pattern.33 The type of exercises that can be evaluated is also limited by the fact that, except in the case of using wireless ultrasound probes, the subject must always be positioned a short distance from the ultrasound cart.33 72 Second, the ultrasound image has a spatial limitation directly related to the length of the ultrasound transducer, especially affecting the measurement of long tendons.72 This limitation could be obviated by scanning only the myotendinous junction.72 However, this requires assuming that the movement of the distal structures to which the tendons attach is negligible, and this does not appear to be the case even with isometric contractions.72 For this reason, it is recommended to scan both tendon ends, using longer transducers when necessary.72 Third, another of the key limitations of ultrasonography is due to the use of 2D images to assess a tendon deformation that occurs in three dimensions.33 72 While the measurement is done through the identification and tracking of anatomical sites in planar 2D images, the reality of 3D movement means that tendon bulging, rotation or twisting can occur, and this fact may introduce a systematic overestimation or underestimation of tendon length.33 72 This limitation has been partially addressed with new 3D ultrasound techniques by capturing images in multiple static postures (eg, Freehand 3D94). In this technique, the ultrasound transducer is moved along the tendon, and a 3D image is created by reconstruction of the captured 2D images. However, this technology requires remaining in a static position for relatively long periods of time to scan the different planes, so its use is limited to resting states or for sustained static contractions.33 72 Some strategies have been suggested to minimise these limitations as much as possible. Some of the most relevant are available in table 1 of the article by Seynnes et al.72

Other techniques

During the review process, other techniques were identified. However, its current application is limited to isometric contractions, exercises such as walking, running or cycling, or controlled contractions in a laboratory setting.

Magnetic resonance imaging

Some authors have used MRI as an imaging technique to measure tendon strain. Finni et al112 in knee extension-flexion cycles against calibrated resistance. Sheehan and Drace92 used phase-contrast cine MRI for evaluating the patellar tendon strain during active knee extensions. In both cases, the reference zero length was identified by analysing MRI images of the tendon in a movie loop of film, noting the joint angle at which the tendon was slack.92 112 This technique allows a 3D analysis, reducing some of the limitations of ultrasonography. However, the nature of the MRI technique makes it difficult to evaluate exercises that require greater mobility.

Stretchable strain sensors

Novel stretchable strain sensors, based on soft elastomers and nanomaterials, are showing great potential for directly measuring musculoskeletal soft tissue strains in vivo.33 These sensors provide direct strain measurement (not force as most of the other available transducers), so they can offer very representative values of the tendon strain. However, these strain sensors share many of their limitations with other implantable devices and must be biotolerable, biocompatible and easy to implant.33

Vibrational behaviour

A proof-of-concept study was identified with a novel technique for evaluating tendon force during walking, running and unilateral and bilateral heel raising.113 Tendon loads were measured using a vibration motor and an accelerometer placed 2 cm apart from each other on the skin superior to the Achilles tendon. The systems consist of exciting a vibration motor and collecting the signals influenced by the tendon force in the accelerometer.113 It is suggested that a tendon on which low force is applied responds to vibration with a steeper rising and falling edge, attributable to faster energy absorption and dissipation.113 However, a tendon on which high force is applied responds with a progressive rising and falling edge, attributable to slower energy absorption and dissipation.113

Another novel non-invasive approach is being developed for in vivo evaluation by tracking vibrational behaviour.114 In this case, the direct relationship between axial stress and the speed of shear wave propagation is exploited through tensiometers consisting of a piezo-actuated tapper and two skin-mounted miniature accelerometers.114

Although these techniques have some limitations such as artefacts caused by noise on the skin caused by movement of the limbs,113 their non-invasiveness gives them an advantage over other evaluation methods.

Limitations

The main limitation of this study is the difficulty in tracking the literature because of the variety and heterogeneity of terms used. This limitation has been minimised through a search including broad terms, but some studies might still not have been identified.

Conclusions

Different evaluation methodologies are used for quantifying tendon forces and strain. However, only a minority of these techniques have been transferred to the study of dynamic rehabilitation exercises. There is a predominant use of modelling and inverse dynamics, but force transducers and optic fibre sensors have also been used for measuring tendon force. Ultrasound imaging is used for measuring tendon strain. Direct force or strain measurement techniques provide significant data, but their current limitations and high invasiveness reduce their application context. Indirect force estimation through inverse dynamics is not invasive but requires making controversial assumptions that may limit its accuracy. Assessing strain using imaging techniques, as long as its limitations are controlled, is a non-invasive method to assess a direct response to the loads acting on the tendon. There are other potentially applicable methods, but they have not yet been transferred to the study of dynamic rehabilitation exercises, possibly due to the difficulty of overcoming some of their limitations.

Although the methods collected in this review allow direct or indirect estimation of the forces and strain applied to the tendon during dynamic exercises, their very nature makes their applicability difficult in a clinical context. Research can use these tools to make general estimates of forces and strain in dynamic exercises, but the invasiveness of some methods and the loss of immediacy of others make it difficult to study each patient individually and provide immediate feedback to the individuals measured. The field should continue to be developed, looking for precise, direct techniques with less measurement error and less invasiveness.

Data availability statement

No data are available. No additional data available.

Ethics statements

Patient consent for publication

Ethics approval

Not applicable.

References

Supplementary materials

  • Supplementary Data

    This web only file has been produced by the BMJ Publishing Group from an electronic file supplied by the author(s) and has not been edited for content.

Footnotes

  • Contributors All authors contributed to the study design. AE-E and AIC-V searched and screened the articles, with assistance from JC. All authors contributed to data analysis and interpretation of the data. AE-E drafted the manuscript; AIC-V and JC revised it critically, and all authors contributed to revisions and approved the final manuscript. Guarantor: AIC-V.

  • Funding This work is part of a government-funded project supported by the University Teaching Training Programme (FPU) of the Ministry of Science, Innovation and Universities of Spain. Grant number: FPU17/00161. The University of Malaga has supported this study with the contribution of funds to support its publication in open access.

  • Competing interests None declared.

  • Patient and public involvement Patients and/or the public were not involved in the design, or conduct, or reporting, or dissemination plans of this research.

  • Provenance and peer review Not commissioned; externally peer reviewed.

  • Supplemental material This content has been supplied by the author(s). It has not been vetted by BMJ Publishing Group Limited (BMJ) and may not have been peer-reviewed. Any opinions or recommendations discussed are solely those of the author(s) and are not endorsed by BMJ. BMJ disclaims all liability and responsibility arising from any reliance placed on the content. Where the content includes any translated material, BMJ does not warrant the accuracy and reliability of the translations (including but not limited to local regulations, clinical guidelines, terminology, drug names and drug dosages), and is not responsible for any error and/or omissions arising from translation and adaptation or otherwise.