Article Text

Download PDFPDF

Original research
Probability of receiving a high cumulative radiation dose and primary clinical indication of CT examinations: a 5-year observational cohort study
  1. Cécile R L P N Jeukens1,
  2. Hub Boere1,
  3. Bart A J M Wagemans1,
  4. Patty J Nelemans2,
  5. Estelle C Nijssen1,3,
  6. Rebecca Smith-Bindman4,5,
  7. Joachim E Wildberger1,3,
  8. Anna M Sailer1,6
  1. 1Department of Radiology and Nuclear Medicine, Maastricht University Medical Centre+, Maastricht, The Netherlands
  2. 2Department of Epidemiology, Maastricht University, Maastricht, The Netherlands
  3. 3CARIM, School for Cardiovascular Diseases, Maastricht University Medical Center, Maastricht, The Netherlands
  4. 4Department of Radiology and Biomedical Imaging Epidemiology and Biostatistics, University of California San Francisco, San Francisco, California, USA
  5. 5Department of Obstetrics Gynecology and Reproductive Sciences, Philip R Lee Institute for Health Policy Studies, University of California, San Francisco, San Francisco, USA
  6. 6Department of Radiology, Stanford University School of Medicine, Stanford, California, USA
  1. Correspondence to Dr Cécile R L P N Jeukens; cecile.jeukens{at}mumc.nl

Abstract

Objective High radiation exposure is a concern because of the association with cancer. The objective was to determine the probability of receiving a high radiation dose from CT (from one or more examinations within a 5-year period) and to assess the clinical context by evaluating clinical indications in the high-dose patient group.

Design Observational cohort study. Effective radiation dose received from one or more CT examinations within a predefined 5-year calendar period was assessed for each patient.

Setting Hospital setting.

Participants All patients undergoing a diagnostic CT examination between July 2013 and July 2018 at the Maastricht University Medical Center.

Primary and secondary outcome measures The primary outcome was the probability of receiving a high effective dose, defined as ≥100 mSv, from one or more CT examinations within 5 years as derived from a time-to-event analysis. Secondary outcomes were the clinical indication for the initial scan of patients receiving a high effective dose.

Results 100 672 CT examinations were performed among 49 978 patients including 482 (1%) who received a high radiation dose. The estimated probability of a high effective dose from a single examination is low (0.002% (95% CI 0.00% to 0.01%)). The 4.5-year probability of receiving a high cumulative effective dose was 1.9% (95% CI 1.6% to 2.2%) for women and 1.5% (95% CI 1.3% to 1.7%) for men. The probability was highest in age categories between 51 and 74 years. A total of 2711 (5.5%) of patients underwent more than six CT examinations, and the probability of receiving a high effective dose was 16%. Among patients who received a high effective dose, most indications (80%) were oncology related.

Conclusions The probability of receiving a high radiation dose from CT examinations is small but not negligible. In the majority (80%) of high effective dose receiving patients, the indication for the initial CT scan was oncology related.

  • computed tomography
  • radiation oncology
  • health policy
http://creativecommons.org/licenses/by-nc/4.0/

This is an open access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited, appropriate credit is given, any changes made indicated, and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/.

Statistics from Altmetric.com

Request Permissions

If you wish to reuse any or all of this article please use the link below which will take you to the Copyright Clearance Center’s RightsLink service. You will be able to get a quick price and instant permission to reuse the content in many different ways.

Strengths and limitations of this study

  • 100 672 CT examinations performed between July 2013 and July 2018 at the Maastricht University Medical Center among 49 978 unique patients were evaluated to estimate the probability of receiving a high dose (≥100 mSv) from one or more CT examinations.

  • Effective doses were calculated for each CT examination separately using a Monte Carlo module allowing to incorporate examination specific scanner settings.

  • A time-to-event analysis was performed which accounts for differences in follow-up period between patients and provides a graph that visualises the increase in probability of receiving a high effective dose over time.

  • The clinical indication for the initial scan of 482 high-dose receiving patients was extracted from medical records.

  • The analysis was restricted to CT examinations performed at a single centre serving as a community hospital but also a tertiary medical centre, which may limit the ability to make generalisations.

Introduction

In the 1970s, the development of CT paved the way for advanced diagnostic imaging. Improvements in technology over time has further increased the use of CT, including the introduction of CT perfusion studies, four-dimensional imaging, dual-energy CT imaging and high-speed CT-angio whole body imaging.1 This increasing clinical value of CT imaging has resulted in an increase in the number of CT examinations over the last decade.2–5 As many diseases and health conditions persist over time, there is also a potential increase in the number of examinations a patient will undergo in his or her lifetime due to monitoring. A drawback of CT imaging is the use of ionising radiation, which for high exposure levels may incur a risk of cancer from as early as 2–5 years following imaging (leukaemia) to a persisting elevated risk that may last many decades (solid cancers).6 7

The risk of cancer associated with radiation exposure levels similar to those used for medical imaging is conventionally assessed using the linear no-threshold model.6 8 9 In this model, even the smallest radiation exposure level is considered to pose a risk, as extensive epidemiological data suggest effective doses in the range of 10–50 mSv are carcinogenic.6 10 Conversely, there remain organisations that question the risks associated with these exposures levels. Some organisations, such as the Health Physics Society and the American Association of Physicists in Medicine,11 12 suggest that cancer risks are small or non-existent with exposure levels below 100 mSv. However, effective doses of 100 mSv and above are consistently considered carcinogenic, with numerous efforts made to reduce the number of individuals exposed to these high effective doses.

In an effort to reduce radiation doses and the associated risks, CT vendors have developed technical solutions including dose modulation and iterative reconstruction methods. As a result, dose reductions have been achieved for some procedures, such as routine chest or abdominal CT where a single phase is obtained. However, technical developments permitting faster scans and use of multiphase studies have also led to imaging that results in higher effective doses. In actual practice, while effective dose reduction has been possible over time, it has often been relinquished in favour of higher effective doses that lead to improved image quality. Furthermore, patients are often imaged repeatedly over time and may be receiving a high cumulative effective dose from multiple CT examinations. There is paucity of information in European settings on the percentage of patients undergoing CT examinations who accumulate high effective doses from recurrent CT examinations and clinical settings in which patients are likely to receive such high effective dose. On the topic of high effective dose prevalence or subsequent clinical settings for the exams, some answers can be found in literature,13–18 none of which originate from our country. Furthermore, the incidence range of patients receiving high cumulative effective doses as reported in literature is wide: from 0% to 15% depending on study design and follow-up period. For some oncology patients with shortened life expectancy, cumulative effective doses may be considered of less importance since low 5-year survival implies low risk of radiation-induced malignancies. However, a recent study by Rehani et al19 found that 14.3% of the group of patients that received high effective doses (>100 mSv) for non-malignant diagnoses were relatively young (<40 years).

To fill the gap in clinical literature on the topic, this observational study was conducted to determine the probability of receiving a high cumulative effective dose (≥100 mSv) from one or more CT examinations and to assess the primary clinical indication for imaging of high effective dose-receiving patients.

Methods

All CT examinations (diagnostic and CT-guided interventions) performed at Maastricht University Medical Center between 1 July 2013 and 1 July 2018 were included, and follow-up was assessed through 31 December 2019. This centre is primarily a tertiary university hospital while also serving as a community hospital. State-of-the-art equipment, advanced protocolling and dose quality assurance are the standard.20–23 The CT scanners included in this study are given in table 1. Follow-up was enabled by a single identification number for each patient.

Table 1

Type of CT scanners, manufacturers, data collection period and distribution of the included CT examinations

Data collection

Radimetrics dose management software (Radimetrics Enterprise Platform, Bayer Healthcare, Berlin, Germany), was used to collect age (at the time of the first examination), sex, date and effective dose of each CT examination. All data were exported to an Excel format file (Microsoft Office Professional Plus 2010, Microsoft Corporation, Redmond, USA) for further analysis.

Effective dose calculation

Effective dose was used to quantify radiation dose for each examination. Effective dose is the sum of the organ doses multiplied by tissue weighting factors, which depend on radio sensitivity of each organ but not on age or sex.8 It is a useful measure as it reflects both the total amount of radiation emitted from the scanner and the future cancer risk associated with those exposures. Thus, the effective dose is a whole body surrogate measure related to the radiation detriment. The unit of effective dose is the sievert, which equals 1 joule per kg (1 Sv=1 J/kg). In diagnostic imaging, effective doses are usually in the millisievert (mSv) range.

Effective doses were calculated using the Monte Carlo-based module embedded in the dose management software.24 To accommodate for patient variations in length and weight, the module incorporates phantom models including 17 different sizes and three pregnant versions. After mapping the CT examination onto the phantom model most closely matching patient characteristics, the effective dose is calculated including incorporation of specific scanner settings.24 Exams with an effective dose of 100 mSv or greater were considered high dose. Patients who received a cumulative effective dose of 100 mSv or greater over the 5 years of the study were considered to have a high dose.

Statistical analysis

The primary analysis estimates the probability of receiving a high cumulative effective dose with a single exam or within the 5-year study period after the initial CT scan. All CT examinations performed on one patient were counted irrespective of indication or examination type. The cumulative effective dose, total observation period and number of CT examinations were calculated for each patient using a custom software script written in Python (Python V.3.4.4, Python Software Foundation, www.python.org). Among patients who received a high cumulative radiation dose at any one time during the study period, the clinical indication for the first CT scan was determined. For these high-dose patients, electronic medical records were individually reviewed with respect to the primary indication for imaging during the follow-up period ending on 31 December 2019.

Nominal and categorical variables are reported as absolute numbers and percentages. Continuous variables are reported as a mean with SDs (if normally distributed), or as a median with IQRs or range (for skewed distributions).

Time-to-event (Kaplan-Meier) analyses were performed to estimate the probability of having the event of interest: a high cumulative effective dose. Follow-up of a patient started at the date of the first CT examination within the 5-year inclusion period (between 1 July 2013 and 1 July 2018). For the primary analysis of probability of receiving a high effective dose, the follow-up ended on the date of which a high cumulative effective dose (≥100 mSv) was reached or observations were censored at the end of the inclusion period (1 July 2018). The number of CT examinations recorded was equal to the number of CT examinations received during the follow-up period. The estimates of probability of receiving a high effective dose were calculated including 95% CIs for the total population and for subgroups according to sex, age and number of CT examinations. Differences between groups were tested using the log rank test. All statistical analyses were performed using SPSS (IBM SPSS Statistics; V.25) and Stata (Survival Analysis, Statacorp, V.13). P<0.05 was considered statistically significant.

Medical ethics approval

The local medical ethics committee approved the study and waived the requirement to obtain informed consent (METC 14-4-158).

Patient and public involvement

There was no formal involvement of patients or members of the public in developing the research questions, design and execution of the study, or interpretation of the results. There are no specific plans to disseminate the results to study patients beyond the usual channels of providing patient information in our hospital.

Results

In total, 104 428 CT examinations were performed within the inclusion period (103 294 diagnostic CT examinations and 1134 CT-guided interventions; figure 1). A total of 3756 CT examinations (3.6%) were excluded because effective dose data were unavailable. These excluded CT examinations were mostly (77%) extremity CT scans (ie, of peripheral body structures such as the hand and joints such as the knee). The remaining 100 672 CT examinations were all included in the analysis (46,710 CT examinations in women and 53 962 in men).

Figure 1

Screening and inclusion profile. MUMC+, Maastricht University Medical Centre+.

The 100 672 included CT examinations were done in a total of 49 978 unique patients (23 918 women and 26 060 men). Median age was 62 (range 0–103) years for women and 62 (range 0–102) years for men. A total of 2130 (4.3%) patients were <18 years old (median 11, range 0–17.9). The distribution of the number of CT examinations per patient is shown in table 2: median (IQR) number of CT examinations per patient was 1 (1–2), the maximum number of CT examinations per patient was 29. The majority of patients, n=39 499 (79%), received one to two CT examinations within the study period. A percentage of 4.7 received 6–10 CT examinations and 0.8% received ≥11 CT examinations within the study period.

Table 2

Distribution of the number of CT examinations per patient in the 5-year study period

Probability of receiving a high (cumulative) effective dose

The distribution of effective doses received in a single examination was highly skewed (figure 2). Only one patient received a high effective dose during a single CT exam, which translates to a probability of 0.002% (95% CI 0.00% to 0.01%).

Figure 2

(A) Histogram of effective doses of 100 672 CT examinations. (B) Enlargement of the histogram to visualise the number of examinations with effective doses >40 mSv.

Kaplan-Meier curves are shown in figure 3, which visualise the probability of receiving a high cumulative effective dose over time according to sex, age and total number of CT examinations. The Kaplan-Meier plots are presented along with the ‘numbers at risk’ below each graph. These numbers indicate how many patients were still in the cohort at that time point. Only a small number of patients had a follow-up period exceeding 4.5 years; therefore, cumulative probabilities at 4.5 (not 5) years after the first CT examination are presented. The median follow-up period (from the date of the first CT examination) was 2.7 years (range 1 day to 5 years).

Figure 3

Kaplan-Meier curves that visualise the probability of receiving a cumulative effective dose of ≥100 mSv over time by (A) sex, (B) age category and (C) number of CT examinations within the study period. The follow-up time starts at the first CT examination.

The probability of receiving a high cumulative effective dose shows a gradual linear increase over time and differed between the categories of sex (figure 3A) and age (figure 3B) and number of CT examinations received (figure 3C).

The 4.5-year probability of receiving a high cumulative effective dose was 1.9% (95% CI 1.6% to 2.2%) for women and 1.5% (95% CI 1.3% to 1.7%) for men; 2.5% for ages 51–64 years (95% CI 2.1% to 2.9%), 2.4% for ages 65–74 years (95% CI 2.0% to 2.9%), 1.1% for ages 18–50 years (95% CI: 0.9% to 1.4%; p≤0.05) and 0% for ages <18 years; 0.01% for patients with 1–2 CT examinations (95% CI 0.00% to 0.03%); 16% for patients with 6–10 CT examinations (95% CI 14% to 18%); and 32% for patients with 11 or more CT examinations (95% CI 26% to 39%).

Baseline characteristics and primary clinical indications for imaging of patients receiving high effective dose

Baseline characteristics of high-dose patients are given in table 3. A total of 482 patients received a high effective dose during the study period. Median age was 63 years (range 20–89). The majority of these patients (52.9%) received 6–10 CT examinations during the study period. The median follow-up period of high-dose patients was 4.2 years (range 12 days–6.5 years).

Table 3

Baseline characteristics of high-dose patients (n=482)

Primary clinical indications for CT examinations in patients receiving high effective doses are given in table 4. The majority of indications (80%) were related to evaluation or follow-up of a malignancy (wide variety of malignancies, predominantly breast cancer, urological and gastrointestinal malignancies), 8% to vascular disease (predominantly aortic aneurysm and dissection) and 8% to non-malignant abdominal disease (predominantly pancreatitis and complicated colitis).

Table 4

Clinical indication for imaging in high-dose patients (n=482)

Discussion

This study included 100 672 CT examinations performed among 49 978 patients at a single tertiary referral centre within a 5-year calendar period. The probability of receiving a high effective dose (≥100 mSv) in a single CT exam was low (0.002% (95% CI 0.00% to 0.01%)). Nearly 40% of patients received multiple CT examinations, and the probability of receiving a high cumulative effective dose after 4.5 years was 1.9% for women and 1.5% for men. The probability of receiving a high effective dose from one to two CT examinations was 0.01% versus 16% from 6 to 10 CT examinations and 32% from ≥11 CT examinations.

Four hundred and eighty-two patients were identified as receiving a high cumulative effective dose, with a median age of 63 years (range 20–89). More than half of these patients underwent 6–10 CT examinations within the study period. As could be expected, almost all high-dose receiving patients underwent more than one CT examination. This is in line with a recent study25 where of a cohort of 8952 patients receiving a high effective dose (≥100 mSv), only 33 received this in a single day or single procedure, which were all CT-guided interventions. Eighty per cent of the CT examination referrals for high-dose patients were oncology related. The latter is similar to findings reported in the literature,13 14 although Rehani et al 19 found 90% oncology-related referrals. At the centre of the current study, continuous effort is undertaken to optimise dose and image quality.20 22 23 While radiation doses at this centre are similar to those in Switzerland and Germany, they were found to be lower than radiation doses in the USA.21

These findings support tracking radiation dose on an individual level. Of course, every exam involving ionising radiation should be properly justified. Patient radiation tracking might hypothetically lead to increased dose awareness among radiologists and referring physicians. In a time where use of CT imaging continues to increase, awareness of the cumulative radiation exposure and the accompanying potential health risk, in particular for younger patients, may lead to more careful and deliberate consideration and justified use of CT examinations.

In the current study, radiation exposure was quantified by calculating the effective dose using a Monte Carlo-based module embedded in the dose management software. The effective dose was introduced by the International Commission on Radiological Protection as a measure for exposure to moderate and low levels of radiation.8 Effective dose can be related to health detriment but was not primarily intended to monitor individual risks. For risk–benefit assessment, the recommended method is to estimate the absorbed organ doses.8

Previous studies have reported percentages of patients receiving high cumulative radiation doses due to diagnostic CT imaging.13–17 However, reported results are difficult to compare with the current study due to differences in study design, population, follow-up and effective dose estimation methods (where either typical effective doses per CT examination13–15 or estimated effective doses16 17 26 were used). Also, a recent study by Rehani and Hauptman18 demonstrated up to a sixfold variation between countries in the estimated number of patients receiving high cumulative effective doses relative to the population of a country. Two population-based studies14 15 used well-defined and large study populations living in specific areas of Spain (224 751 people)14 and the USA (54 447 adults).15 They reported analyses of the subgroups of inhabitants undergoing at least one CT examination in the study period. Reported estimates of the percentage of patients receiving a cumulative effective dose >100 mSv were 8.2% for a 12-year period in Spain and 1.9% for a 10-year period in the USA. The authors of the latter study indicate that they likely underestimated cumulative effective doses and percentages because average effective dose estimates per CT examination were taken. This does not account for higher BMI patients receiving higher effective doses. Conversely, the Spanish study is likely an overestimation as exposure was extrapolated, assuming the same exposure rate for patients that could not be followed for 12 years.

A few hospital-based studies report percentages of patients reaching a cumulative effective dose ≥100 mSv in patients referred for a CT examination. The highest percentage was reported by Sodickson et al13 in a cohort of 31 462 patients undergoing CT in a tertiary academic medical centre in the USA. Evaluation of all recorded CT exams over 22 years leading up to 2007 showed that 15% of patients received an effective dose ≥100 mSv. The percentage of patients receiving more than five CT examinations in this study is high (33%/22 years), which may explain the relatively high incidence of high-dose patients found. In a study by Rehani et al,16 in patients from four centres (three in USA and one in Central Europe), the percentage of patients receiving a high effective dose ranged from 0.64% to 3.4% (average 1.3%) for a follow-up duration of 1–5 years. A survey yielding data from 20 countries17 reported percentages ranging from 0.0% to 5.0% with a follow-up duration ranging from 0.4 to 6.1 years. None of these hospital-based studies performed time-to-event (survival) analyses to account for differences in varying follow-up periods between patients, which might have led to bias.27 28

A strength of the current study is that effective doses were calculated for each CT examination separately using a Monte Carlo-based module. This is in line with previous research16 17 19 and has the advantage of incorporating scanner-specific settings such as scan range rather than using generic literature-based effective dose values. Another strength of the current study is the Kaplan-Meier analysis, which accounts for differences in follow-up periods between patients and provides a graph that visualises the increase in probability of receiving a high effective dose over time. Finally, the current paper adds information on the European context to the mainly US-centric information in current literature. Studies by Smith-Bindman et al show that effective doses at US sites are generally higher than those at European sites.21

There are some limitations to our study. First, the analysis on cumulative effective dose was restricted to CT examinations that were performed at a single centre, which limits the ability to generalise. Maastricht University Medical Center serves as community hospital and a tertiary medical centre. The probability of receiving a high cumulative effctive dose for patients who are referred to a tertiary centre will strongly depend on efforts undertaken to optimise the use of state-of-the-art CT scanners with respect to dose and image quality. CT examinations performed at other centres were not included in the cumulative effective dose analysis, nor was radiation exposure due to other sources such as conventional X-ray, fluoroscopy and nuclear medicine. As a result, cumulative effective doses will tend to underestimate. A potential solution for this matter requires a unique registration of patients over time in a broad perspective. Dose-related data can be tracked based on a unique personal number like the BSN in the Netherlands. This will require political support on that matter on a national or even a European level. Second, a recent study29 has shown that the absorption of contrast medium can considerably increase the absorbed dose in contrast-uptaking organs. The Monte Carlo-based module embedded in the dose management software used to estimate the effective dose in the current study does not incorporate this effect and may underestimate effective dose in contrast to media CT examinations. Third, the time-to-event analysis could only partially account for differences in follow-up period between patients. In the primary analysis, censoring the observations of patients who died or were referred to another hospital before the end of the inclusion period was not possible, because this information was not available. As a result, the numbers at risk during follow-up are overestimated and the probability of receiving a high cumulative effective dose may have been underestimated. The aforementioned limitations result in underestimation of cumulative effective dose, and therefore results must be interpreted as being minimum probabilities, further underscoring the necessity of tracking radiation dose.

Conclusions

The probability of receiving a high cumulative effective dose ≥100 mSv from CT examinations is small but not negligible and increases linearly with time. In the majority (80%) of high-dose-receiving patients the indication for the initial CT scan was oncology related. These findings advocate the need for patient dose tracking to gain insight into individual cumulative radiation exposure and to increase awareness of the radiation exposure among radiologists and referring physicians.

References

Footnotes

  • Correction notice This article has been corrected since it was published. Affiliation 3 has been added to Joachim E Wildberger.

  • Contributors CRLPNJ, JW, PJN and AMS designed the study. CRLPNJ, HB, BAJMW and AMS collected the data. CRLPNJ, HB, PJN and AMS analysed the data. CRLPNJ, PJN, RSB, JW and AMS interpreted the data. CRLPNJ, ECN and AMS drafted the manuscript. All authors critically evaluated, revised and approved the manuscript. The corresponding author attests that all listed authors meet authorship criteria and that no others meeting the criteria have been omitted. CRLPNJ is the guarantor.

  • Funding The authors have not declared a specific grant for this research from any funding agency in the public, commercial or not-for-profit sectors.

  • Competing interests None declared.

  • Patient and public involvement Patients and/or the public were not involved in the design, or conduct, or reporting, or dissemination plans of this research.

  • Patient consent for publication Not required.

  • Provenance and peer review Not commissioned; externally peer reviewed.

  • Data availability statement Data are available on reasonable request. On request, the used Python code or anonymised raw data can be shared by contacting the corresponding author.