Table 2

Diagnostic performance of WatchBP and Omron autoanalysis for detecting atrial fibrillation

n=999* Prevalence 7.9%WatchBP with AF indicator flash (95% CI)Omron autoanalysis* (95% CI)
Sensitivity (%)94.9 (87.5 to 98.6)98.7 (93.2 to100)
Specificity (%)89.7 (87.5 to 91.6)76.2 (73.3 to 78.9)
Positive likelihood ratio9.2 (7.6 to 11.2)4.15 (3.69 to 4.67)
Negative likelihood ratio0.057 (0.022 to 0.15)0.017 (0.0024 to 0.12)
Positive predictive value44.1 (36.5 to 51.9)26.3 (21.3 to 31.7)
Negative predictive value99.5 (98.8 to 99.9)99.9 (99.2 to 100)
  • *Text message ‘irregular’ or ‘analysis impossible’ counted as positive test.

  • AF, atrial fibrillation.