Table 2

Number of attempts of intubation trial for successful endotracheal intubation in three scenarios

Number of attempts for success
DevicesNumber of successful ETIs1 (%)2 (%)3 (%)p Value
Manual in-line stabilisation
 Direct laryngoscope3330 (90.9)2 (6.1)1 (3.0)0.795
 GlideScope3230 (93.8)1 (3.1)1 (3.1)
 PAWS3534 (97.1)1 (2.9)0 (0.0)
Neck immobilisation with a Philadelphia neck collar
 Direct laryngoscope2524 (96.0)1 (4.0)0 (0.0)0.044
 GlideScope3328 (84.8)5 (15.2)0 (0.0)
 PAWS2927 (93.1)0 (0.0)2 (6.9)
Neck immobilisation with a Stifneck collar
 Direct laryngoscope1413 (92.9)1 (7.1)0 (0.0)0.627
 GlideScope2424 (88.9)3 (11.1)0 (0.0)
 PAWS77 (100)0 (0.0)0 (0.0)
  • ETI, endotracheal intubation; PAWS, Pentax Airway Scope.