A pilot study of a wearable apnoea detection device
  1. Esther Rodriguez-Villegas1,
  2. Gwangwei Chen1,
  3. Jeremy Radcliffe2,
  4. John Duncan3
  1. 1Department of Electrical and Electronic Engineering, Imperial College London, London, UK
  2. 2Department of Neuroanaesthesia, UCL Institute of Neurology, National Hospital for Neurology and Neurosurgery, London, UK
  3. 3Department of Clinical and Experimental Epilepsy, UCL Institute of Neurology, National Hospital for Neurology and Neurosurgery, London, UK
  1. Correspondence to Dr Esther Rodriguez-Villegas; e.rodriguez{at}imperial.ac.uk

Abstract

Rationale Current techniques for monitoring patients for apnoea suffer from significant limitations. These include insufficient availability to meet diagnostic needs, cost, accuracy of results in the presence of artefacts and difficulty of use in unsupervised conditions.

Objectives We created and clinically tested a novel miniature medical device that targets overcoming these limitations.

Methods We studied 20 healthy control participants and 10 patients who had been referred for sleep apnoea diagnosis. The performances of the new system and also of the Food and Drug Administration (FDA) approved SOMNO clinical system, conventionally used for sleep apnoea diagnosis were evaluated under the same conditions. Both systems were tested during a normal night of sleep in controls and patients. Their performances were quantified in terms of detection of apnoea and hypopnoea in individual 10 s epochs, which were compared with scoring of signals by a blinded clinician.

Main results For spontaneous apnoeas during natural sleep and considering the clinician scorer as the gold standard, the new wearable apnoea detection device had 88.6% sensitivity and 99.6% specificity. In comparison the SOMNO system had 14.3% sensitivity and 99.3% specificity. The novel device had been specifically designed to detect apnoea, but if apnoea and hypopnoea during sleep were both considered in the assessment, the sensitivity and specificity were 77.1% and 99.7%, respectively, versus 54% and 98.5%, respectively, for the SOMNO.

Conclusions The performance of the novel device compares very well to the scoring by an experienced clinician even in the presence of breathing artefacts, in this small pilot study. This can potentially make it a real solution for apnoea home monitoring.

  • RESPIRATORY MEDICINE (see Thoracic Medicine)
  • SLEEP MEDICINE

This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/